Figure 2 : Alarm Indicators
4.
Check function of alarms and indicators
(refer to Figure 2)
i)
Set the Function Selector to ‘0’.
ii)
Without a gas supply connected to the ventilator, select CMV + PEEP.
iii)
The gas supply indicators will stay red but the alarm LEDs will flash in sequence, starting
bottom left and rotating in a clockwise direction. After all 6 LEDs have flashed a Medium
Priority alarm will sound and continue for 1 minute. If the ventilator is still left switched on
the alarm system will shut down.
iv)
With the air and oxygen input hoses connected to the ventilator, connect the hose probes to
appropriate gas supplies (minimum requirement 20L/min at 305kPa
≈ 3
bar).
v)
Connect a C0.5 R100 ISO Test Lung (500 - 84250) to the Patient Connection Port of the
Patient Circuit.
vi)
Turn on the air supply and the lower supply gas failure visual alarm should change from red to
quartered white and black. Repeat for the oxygen supply; the upper alarm turning from red to
white.
vii) Set both the Oxygen Concentration and the PEEP/CPAP controls to minimum.
viii)
Select CMV + PEEP on the Function Selector switch. The ventilator should now begin to
cycle and the alarm LEDs will go through the test sequence described in iii). The yellow
‘silenced’ LED (bottom right) should flash but no sound should be generated for 60 seconds.
The ‘cycle’ LED should flash for 1/10 second each time the pressure rises through 10
x100Pa (10 cmH2O).
ix)
Turn off the air supply. As the supply depletes, the air supply gas visual alarm should change
to red. A medium priority electronic audible alarm giving a burst of three pulses 6 times a
minute should then sound for 60 seconds to indicate a supply failure and the ‘single gas
operation’ LED should illuminate in bursts of three flashes every 30 seconds.
Cycle
High inflation
pressure/ Constant
positive pressure
Low battery
Single gas operation/
Supply gas failure
Low inflation pressure
(disconnect)
Silence
button
Electronic alarm
silenced/muted
Summary of Contents for Pneupac babyPAC 100
Page 2: ...504 2038NUS March 2005 Page Intentionally Blank ...
Page 8: ...504 2038NUS 8 March 2005 Page Intentionally Blank ...
Page 20: ...504 2038NUS 20 March 2005 ...
Page 23: ...504 2038NUS 23 March 2005 Page Intentionally Blank ...
Page 27: ...504 2038NUS 27 March 2005 Page Intentionally Blank ...
Page 31: ...504 2038NUS 31 March 2005 ...