Chapter 6: Alarms
Home Use Operation Manual
6-3
High Priority Alarms
A high priority alarm warns of an abnormal patient condition.
A high priority alarm is activated when:
The patient’s SpO
•
2
reading matches or exceeds the SpO
2
alarm range.
The patient’s pulse rate reading matches or exceeds the pulse rate alarm range.
•
There is a lost pulse condition (the sensor no longer detects a pulse while a finger is
•
inserted in the sensor, when a pulse was previously detected).
During a high priority alarm:
The high priority alarm tone sounds, if not silenced. The high priority alarm tone consists of
•
two bursts of five tones each, repeated every 7 seconds.
The numeric display of each exceeded parameter flashes twice per second.
•
During a lost pulse condition, pulse rate and SpO
2
displays will flash “
--
”.
The Alarm Indicator flashes red twice per second.
•
Each time new trend data is saved it is marked with an alarm active flag.
•
The alarm condition is sent to the data output port.
•
Alarms active at the time of a real time data output are indicated on the optional WW1026
•
thermal printer.
nOTe: both the SpO
2
and pulse rate numbers will flash if both readings are equal to or
exceed their alarm range.
nOTe: The alarm actions occur for each violated alarm, even if more than one alarm is
violated at the same time.
Summary of Contents for Oxi-Pulse 30
Page 2: ......
Page 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Page 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Page 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Page 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Page 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Page 65: ......