Alarms
Tips/reason
Operation
Occlusion Downstream blockage
Close roller clamp and press & hold
Audio Pause/Alarm reset key for 2
seconds to clear the alarm. After the
occlusion is eliminated, open the roller
clamp and press the Start key
Air in line Air detected
Press and hold the Audio Pause/Alarm
reset key for 2 seconds to clear
the alarm. Disconnect from patient
and remove the air from tubing or load
the tubing correctly to clear the alarm.
Press Start key
Battery
low
The battery capacity is
low
Press the Mute key , connect the
pump to an AC power supply as soon
as possible
Battery
over
The battery capacity is
empty, the infusion pump
has stopped
Connect the infusion pump to an AC
power supply immediately
Inf.
Ending
The infusion is
about to complete
in 3 minutes
Press the Audio Pause/Alarm reset
and the alarm will be muted for 2
minutes
Inf. Comp
The infusion is
completed, the
pump will operate
in the KVO rate
Press the Audio Pause/Alarm reset
and the alarm or press the Stop key
to stop the infusion
No
Operation
Indicates the pump
has been left idle for 2
minutes
Press any key to silence the alarm
and start the infusion or turn off the
infusion pump
Door
Open
If the lock handle is
open during infusion or
closed incompletely when
the pump starts, it will
trigger an audible and
visual alarm. At the same
time, the pump stops
running or the infusion
cannot be started
The alarm can be cleared by closing
the lock handle. Start key to
continue infusion
Priming
Stuck
Priming volume has
reached 6 mL
Release the Priming /Bolus key
PRIMING
/TUBING
to clear alarm
Alarms
PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN ALL GEOGRAPHIES
Smiths Medical ASD, Inc.
6000 Nathan Lane North
Minneapolis, MN 55442, USA
Tel: 1-614-210-7300
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com
Smiths Medical International Ltd
1500 Eureka Park, Lower Pemberton
Ashford Kent, TN25 4BF
Tel: +44 (0)845 850 0445
www.smiths-medical.com
If the infusion pump alarms, listen for the beeps and check the display
screen for alarm messages or symbols.
Alarm indicator
External power
indicator
Door bolt
Pressure
sensor
Fluid clamp
Ultrasonic
sensor
Dynamic box
Read the entire Instruction Manual before operating the Graseby™
1200 Infusion Pump . Failure to properly follow warnings, cautions and
instructions could result in death or serious injury to the patient.
The Graseby™ 1200 is a volumetric infusion pump that provides accurate
and reliable infusions at a rate of 1 - 1200 mL/h
IS0208.GB.UK.Rev.A.0420.indd 6
IS0208.GB.UK.Rev.A.0420.indd 6
23/04/2020 11:03
23/04/2020 11:03