Alarms

Tips/reason

Operation

Occlusion Downstream blockage

Close roller clamp and press & hold 

Audio Pause/Alarm reset key    for 2 

seconds to clear the alarm. After the 

occlusion is eliminated, open the roller 

clamp and press the Start key 

Air in line Air detected

Press and hold the Audio Pause/Alarm 

reset key    for 2 seconds to clear 

the alarm. Disconnect from patient 

and remove the air from tubing or load 

the tubing correctly to clear the alarm. 

Press Start key 

Battery 

low

The battery capacity is 

low

Press the Mute key , connect the 

pump to an AC power supply as soon 

as possible

Battery 

over

The battery capacity is 

empty, the infusion pump 

has stopped

Connect the infusion pump to an AC 

power supply immediately

Inf. 

Ending

The infusion is 

about to complete 

in 3 minutes

Press the Audio Pause/Alarm reset  

 and the alarm will be muted for 2 

minutes

Inf. Comp

The infusion is 

completed, the 

pump will operate 

in the KVO rate

Press the Audio Pause/Alarm reset    

and the alarm  or press the Stop key

to stop the infusion

No 

Operation

Indicates the pump 

has been left idle for 2 

minutes

Press any key to silence the alarm 

and start the infusion or turn off the 

infusion pump

Door 

Open

If the lock handle is 

open during infusion or 

closed incompletely when 

the pump starts, it will 

trigger an audible and 

visual alarm. At the same 

time, the pump stops 

running or the infusion 

cannot be started

The alarm can be cleared by closing 

the lock handle. Start key   to 

continue infusion

Priming 

Stuck

Priming volume has 

reached 6 mL

Release the Priming /Bolus key 

PRIMING

/TUBING

to clear alarm

Graseby

 1200 

Large Volume Infusion Pump

Quick Reference Guide

PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN ALL GEOGRAPHIES

Infusion tubing brand

Infusion status indicator

Main screen

Time display

Battery indicator

Pressure Indicator

Audio pause/ 

Alarm reset key

Alarm indicator

Stop key

Start key

Power key

External power

 indicator

Select key
Priming/Bolus 

key

Barrel clamp

Pump door

Move cursor key

Shift cursor key

Digit setting key (Up)

Digit setting key (Down)

Door bolt

Infusion 

direction

Pressure 

sensor

Fluid clamp

Ultrasonic 

sensor

Dynamic box

Drag hook

Lock handle

Read the entire Instruction Manual before operating the Graseby™ 

1200 Infusion Pump . Failure to properly follow warnings, cautions and 

instructions could result in death or serious injury to the patient.
The Graseby™ 1200 is a volumetric infusion pump that provides accurate 

and reliable infusions at a rate of 1 - 1200 mL/h

IS0208.GB.UK.Rev.A.0420.indd   1

IS0208.GB.UK.Rev.A.0420.indd   1

23/04/2020   11:03

23/04/2020   11:03

Summary of Contents for Graseby 1200

Page 1: ...Pump door Move cursor key Shift cursor key Digit setting key Up Digit setting key Down Door bolt Infusion direction Pressure sensor Fluid clamp Ultrasonic sensor Dynamic box Drag hook Lock handle Read the entire Instruction Manual before operating the Graseby 1200 Infusion Pump Failure to properly follow warnings cautions and instructions could result in death or serious injury to the patient The ...

Page 2: ...the patient While priming ensure that the tubing is not connected to the patient 1 Open the roller clamp between the patient and the infusion pump 2 Press the Priming Bolus key PRIMING TUBING PRIMING TUBING twice and hold at the second time to prime the infusion tube at the rate of 1000 mL h 3 Release the Priming Bolus key PRIMING TUBING PRIMING TUBING when you observe that the air has been expell...

Page 3: ...sonic sensor slot 2 3 Close the lock handle and press the lock handle in and down to close the pump door Ensure the lock handle is flush with the pump s outer case Not to be used with blood and blood products Load the Administration Set Pumping segment Fluid clamp Flow regulator Lock handle Direction of flow NOTE Ensure the infusion tubing has no fluid leakage after infusion starts When the pump h...

Page 4: ... Infusion rate unit Dynamic drop symbol Infusion rate value Infusion time left Figure 1 Pump screensaver The Graseby 1200 infusion pump can be programmed to infuse in two ways 1 Set the infusion volume and rate and the infusion time will automatically be calculated 2 Set the infusion volume and infusion time and the infusion rate will automatically be calculated Set and Program an Infusion 1 Use t...

Page 5: ...usion Program 1 to set the infusion rate and infusion volume for Infusion Program 2 4 After the correct infusion program is displayed press to move the cursor to OK then press the Select key to confirm and save the infusion program setting and return to the main screen 5 On the Infusion Program Setting screen press and then concurrently to clear the infusion program 6 Press the start key and the p...

Page 6: ... capacity is empty the infusion pump has stopped Connect the infusion pump to an AC power supply immediately Inf Ending The infusion is about to complete in 3 minutes Press the Audio Pause Alarm reset and the alarm will be muted for 2 minutes Inf Comp The infusion is completed the pump will operate in the KVO rate Press the Audio Pause Alarm reset and the alarm or press the Stop key to stop the in...

Reviews: