Smiths Medical CADD-Legacy PLUS Quick Reference Card Download Page 2

PRoGRAmmInG / SET-UP SEQUEnCE

ConTInUoUS DELIVERY

Batteries must be installed; pump must be stopped and in LL1.

noTE: Value not Saved is displayed if a value is scrolled and EnTER/

CLEAR is not pressed. Press nEXT to continue programming.
WARnInG: If the pump is dropped or hit, inspect it for damage. Do not 

use a pump that is damaged or is not functioning properly.

View Dose Cycle

PUmP oPERATIonS

1.  Stop the pump.
2.  Press LOCK.
3.  View lock level.
4.  Press LOCK or ENTER/CLEAR two more times.

1. Press and hold STOP/START until  

(– – – –  – – – –  – – – –) appears on the display.

2. Release STOP/START key. STOPPED will appear  

on the display when the pump is stopped.

1.  Press and hold STOP/START until  

(– – – –  – – – –  – – – –) disappears from the display. 

2.  Release STOP/START key. RUN will appear on the 

display when the pump is running.

Pump must be stopped and in LL0 or LL1.

Warning: Do not prime the fluid path with the  

tubing connected to a patient as this could result in  

over-delivery of medication or air embolism.

1.  Press and hold PRIME until the word PRIME 

appears  on  the  display,  along  with  (– – – –  – – – –  – – – –).

2.  Release the PRIME key.
3.  Press and hold PRIME. Priming appears on the 

display. Continue priming until the fluid path is  
free of air.

4.  Press NEXT to return to the main screen.

1. Stop the pump.
2. Press NEXT to display the reservoir volume screen.
3. Press ENTER/CLEAR to reset the value to previously 

programmed amount.

1. Press and hold ON/OFF until pump beeps  

and powers up.

1. Press and hold ON/OFF until (•••• •••• ••••) 

appears on the display.

2. Release ON/OFF key.
3. The screen will go blank as the pump goes into  

a lower power state.

1. Stop the pump.
2. Push down and hold the arrow button on the  

battery door while sliding the door off.  
Remove and discard old batteries.

ImPoRTAnT: Always stop pump before  

removing batteries.

3. Install new batteries, matching polarities shown  

on pump. Replace battery door and close. 

4. Start the pump.

Warning: If a gap is present anywhere between the 

battery door and the pump housing, the door is not 

properly latched. If the battery door becomes  

detached or loose, the batteries will not be properly 

secured which could result in loss of power or  

non-delivery of medication.
Warning: Do not use rechargeable niCad or nickel  

metal hydride (nimH) batteries. Do not use carbon  

zinc (“heavy duty”) batteries. 

View Reservoir 

Volume
Enter Continuous 

Rate

Clear milliliters 

Given

Verify Air  

Detector Status
Verify Upstream   

Sensor Status

Verify  

Programming

main Screen

View Reservoir 

Volume
View Dose Volume

View Dose Duration

main Screen

View KVo Rate

Verify Dose  

Remaining

View the 

Lock 

Level

Stop the 

Pump

Start the 

Pump

Press NEXT.

1.  Press ENTER/CLEAR to reset.
2.  Press NEXT.

3. Press

´

or

Î

to select desired Continuous   

Rate in ml/hr.

4.  Press ENTER/CLEAR.
5.  Press NEXT.

Warning: This pump delivers medication at a  

continuous rate or intermittent rate in milliliters 

per hour (ml/hr). Programming the pump at a 

delivery rate other than what is prescribed will 

cause over- or under-delivery of medication.

6. Press ENTER/CLEAR to clear the given 

value. The display will show 0.00.

7. Press NEXT.

8. Verify the setting is correct.
9. Press NEXT.

10.Verify the setting is correct.
11.Press NEXT.

12.Press NEXT repeatedly to review program.

InTERmITTEnT DELIVERY

Batteries must be installed; pump must be stopped and in LL1.

Press NEXT.

1.  Press ENTER/CLEAR to reset.
2.  Press NEXT.

3.  Press NEXT.

4.  Press NEXT.

5.  Press NEXT.

6.  Press NEXT.

note: Appears only during dosing or if dose is 

interrupted.

7.  Press NEXT.

8. Press

´

or

Î

 to select amount of time 

before next dose. Select IMMEDIATE if 
dose should begin when pump is started.

9.  Press ENTER/CLEAR.
10. Press NEXT.

11.  Press ENTER/CLEAR to clear the given  

value. The display will show 0.00.

12. Press NEXT.

13. Verify the setting is correct.
14. Press NEXT.

15. Verify the setting is correct.
16. Press NEXT.

17. Press NEXT repeatedly to review program.

Enter Dose

Starts In

Clear milliliters 

Given

Prime the 

Fluid Path

Reset the 

Reservoir 

Volume

Turn the 

Pump on

Turn the 

Pump off

Change the 

Batteries

Verify Air Detector 

Status
Verify Upstream  

Sensor Status

Verify 

Programming

PRoGRAmmInG SCREEnS

main Screen 

Displays status of pump  

 

 

(STOPPED or RUN)

ConTInUoUS moDE

Reservoir Volume

 

Volume of fluid in reservoir

Continuous Rate

 

Continuous rate of infusion (ml/hr)

Given

 

 

Total amount of medication delivered since  

 

 

display was last cleared

Air Detector

 

Indicates if air detector is turned on or off  

(status only)

 

and high or low sensitivity level 

Upstream Sensor

  

Indicates if turned on or off 

(status only)

 

InTERmITTEnT moDE

Reservoir Volume

 

Volume of fluid in reservoir

Dose Volume

 

Volume of an individual dose

Dose Duration

 

Amount of time over which an individual  

 

 

dose is infused

Dose Cycle 

 

Time from start of one dose to start  

 

 

of next dose

KVo Rate

   

Keep Vein Open rate in ml/hr delivered  

  between 

doses 

Dose Remaining

 

Time remaining for delivery of current  

 

 

dose (only appears during dosing or  

 

 

if dose is interrupted)

Dose Starts In

 

Amount of time before start of next dose

Given

 

 

Total amount of medication delivered since  

 

 

display was last cleared

Air Detector

  

Indicates if air detector is turned on or 

(status only)

 

off and high or low sensitivity level 

Upstream Sensor

  

Indicates if turned on or off  

(status only)

IN19893.indd   2