background image

acapella

®

 choice Vibratory PEP System

Reference Guide

These instructions contain important information 
for safe use of the product. Read the entire contents 
of the Instructions for Use, including Warnings and 
Cautions, before using this product. Failure to properly 
follow warnings, cautions and instructions could 

result in death or serious injury to the patient. It is the 
responsibility of the healthcare practitioner to assure 
that the instructions for use and maintenance are 
understood by and provided to the caregiver.

B.

A.

D.

C.

Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

90°

Rib

Figure 6

1.  DESCRIPTION:

The acapella

®

 choice vibratory PEP system is a  

single patient use device that provides Positive 
Expiratory Pressure (PEP) therapy for patients 
who have Cystic Fibrosis, COPD, asthma, and lung 
diseases with secretory problems, and patients 
with atelectasis. All patients must be capable of 
following instructions for PEP therapy. Review the 
diagram and the product to become familiar with 
all of the product features. 
The acapella

®

 choice Vibratory PEP system with  

one-way inspiratory valve and 22 mm male fittings 
consists of: (See Figure 1)

A. 

Detachable mouthpiece  
(device end adapts to mask)

B. 

Expiratory resistance/frequency adjustment dial

C. 

Detachable cover

D. 

Detachable rocker assembly – DO NOT  
disassemble the rocker arm from its  
platform base

2.  INDICATIONS:

The Smiths Medical acapella

®

 choice system is 

intended for use as a PEP device. It may also be 
used simultaneously with nebulized aerosol 
drug delivery.

3.  CONTRAINDICATIONS:

Although no absolute contraindications to the use 
of PEP therapy have been reported, the following 
should be carefully evaluated before a decision is 
made to initiate therapy:

• 

Inability to tolerate increased work of breathing

• 

Hemodynamic instability

• 

Intracranial pressure (ICP) > 20 mm Hg

• 

Acute sinusitis

• 

Recent facial, oral or skull surgery or trauma

• 

Epistaxis

• 

Esophageal surgery

• 

Active hemoptysis

• 

Untreated pneumothorax

• 

Nausea

• 

Known or suspected tympanic membrane rupture  
or other middle ear pathology

4.  WARNINGS:

4.1 

Use of this device at excessive pressures may 
have adverse effects. Expiratory pressures above 
20 cm H2O in patients sensitive to increased 
transpulmonary pressure may develop one or 
more of the adverse side effects listed below.

4.2 

Expert clinical judgment should be exercised in 
the selection of the appropriate setting for each 
individual patient. Failure to match the appropriate 
resistance setting on the +/- dial indicator with 
the patient’s expiratory flow may result in failure 
to achieve therapeutic objectives of vibratory PEP 
therapy or one or more adverse side effects below.

• 

Adverse reactions may include:

• 

Increased work of breathing that may lead to 
hypoventilation and hypercarbia

• 

Increased cranial pressure

• 

Cardiovascular compromise

• 

Myocardial ischemia

• 

Decreased venous return

• 

Air swallowing with increased likelihood of 
vomiting and aspiration

• 

Claustrophobia

• 

Skin break down and discomfort from mask

• 

Pulmonary barotraumas

4.3 

The rocker assembly is not designed to be 
disassembled - do not attempt to separate  
the rocker arm from its platform base. Separating 
these components may cause the acapella

®

  

choice system to function improperly, which may  
deprive the patient of therapy and cause injury to 
the patient.

5.  PRECAUTIONS:

5.1 

Bleach is not recommended for use on the 
acapella

®

 choice system. It may deteriorate the 

nickel plated mechanism located in the interior  
of the device.

5.2 

DO NOT MICROWAVE. The metal and magnet  
might ignite.

5.3 

It is the responsibility of the user to ensure all 
sterility verification(s).

5.4 

Visually inspect the device to ensure that the  
unit is free of contamination and foreign objects.

5.5 

Verify all connections are secure.

Reviews: