20
Functional Testing
The Power‑up Check, Latch/Lock Check
and Cassette Sensor Check using a CADD
™
medication cassette reservoir or CADD
®
administration set can be performed with the
pump in any of the four delivery modes.
Power-up Check
• Insert a battery in the pump and observe the
LCD during power up.
The number values may vary depending on the
current mode and software revision.
• If “lec XXXXX” (last error code with 5 digit
number) appears on the display prior to the
pump reviewing the current program settings,
the pump has experienced an electrical or
mechanical fault and should be returned for
service. If no error message is immediately
shown, the pump has powered up normally.
The pump should sequentially display all of
the programmed values. The words “Self
Test Complete” should appear, then the text
“Power Up Successful” with six audible beeps.
Continue with the Latch/Lock check.
Latch/Lock Check
• Attach a 50‑ or 100‑ml CADD
™
medication cassette reservoir or a CADD
®
administration set to the pump. The mark
on the latch should be aligned with the
solid dot. The display should show that the
reservoir or administration set is latched.
See Table 13.
• Lock the device by inserting a key into the
lock and turning counterclockwise until the
mark lines up with the solid dot. The display
should show “Cassette Locked.”
• Unlock the device by inserting a key into
the lock and turning clockwise until the
mark lines up with the open dot. The display
should show “Cassette Unlocked.”
• Unlatch the reservoir by inserting a coin into
the latch slot and turning clockwise until the
9 Testing Procedures
C A D D - P R I Z M
61XX
lec XXXXX
sn XXXXXXXX
0351-01X
XXXX-X
Software revision level
Error code
Pump model number
Serial number
If the Epidural mode is NOT turned on in the
Options Menu
Reservoir Latched
VIEW to continue
Admin set Latched
Upstream Sensor On.
VIEW to continue
Table 13.
If the Epidural mode is turned on in the
Options Menu
Reservoir Latched
Is epidural reservoir
attached
Press YES or NO
Admin set Latched
Upstream Sensor On.
Is epidural admin set
attached?
<Press YES or NO>
• Select Yes
mark lines up with the open dot. The display
should show “Cassette Unlatched / Close
Clamp to Prevent Free Flow.”
Cassette Sensor Check
• Attach a 50‑ or 100‑ml CADD
™
medication
cassette reservoir to the pump. Latch the
cassette to the pump. The display should
show “Reservoir latched.”
NOTE
The message displayed depends on
the type of reservoir or administration set
attached. See Table 13.
• Lock the device by inserting a key into the
lock and turning counterclockwise until the
mark lines up with the solid dot. The display
should show “Cassette Locked.”
• Unlock the cassette. The display should show
“Cassette Unlocked.” Unlatch the cassette.
The display should show “Cassette Unlatched /
Close Clamp To Prevent Free Flow.”
• Remove the 50‑ or 100‑ml reservoir and
attach a CADD
®
administration set to the
pump. Latch the cassette to the pump. The
display should show “Admin Set Latched.”
(See table 13)
• Lock the device by inserting a key into the
lock and turning counterclockwise until the
mark lines up with the solid dot. The display
should show “Cassette Locked.”