LogiCal
®
5ml Closed Blood
Sampling System
Quick Blood Sampling Operation Guide
•
Verify the reservoir stopcock “OFF”
arrow points to the reservoir syringe
and reservoir syringe plunger is on
0ml position.
Prior to Sampling
1
•
Turn the reservoir stopcock “OFF”
arrow to point upstream to the fl uid
bag.
•
Locate the in-line needle-free
blood sampling site and cleanse the
sampling site per hospital policy
and procedure.
•
Slowly pull the reservoir syringe
plunger down to the 5ml position to
draw back fl uid, bringing undiluted
blood to the sampling site.
•
Turn the reservoir stopcock “OFF”
arrow to point downstream to the
patient.
Prepare the system to obtain
blood samples:
2
•
Connect a Luer lock syringe or
sampling device to the in-line
needle-free blood sampling site.
•
Gently withdraw the desired blood
sample.
•
Disconnect the Luer lock syringe
or sampling device from the in-line
needle-free blood sampling site.
Obtaining blood samples:
3
STEPS 4-5
CONTINUED ON REVERSE
BEFORE YOU START
Remove the kit from the pack, confi rm
all Luer connectors are tightened.
Attach and mount the LogiCal
®
5ml closed blood sampling system
properly to the mounting system.*
Before you start to obtain a blood
sample, verify the system is primed
per the IFU and the fl uid bag is
pressurised to 300mmHg and placed
at a higher level than the system.
* The mounting system consists of the LogiCal
®
transducer plates, clamp, brackets and cables.
See LogiCal
®
5 ml Closed Blood Sampling
System brochure for references.