Securing the stem
Blocks
5 mm,
10 mm and 15 mm thick ( 15 mm
distal only ) are available for the femoral components
( page 6
1
and 6
2
).
Important
When using blocks note that the femoral implants must
be clean and dry. Note the corresponding size
combinations.
The femoral blocks are also fixed mechanically
with the premounted retaining screws with the
screwdriver with torque.
The required torque is
reached when the line reaches the Stem/Block position
( 4.5 Nm ).
Important
The posterior femoral blocks are the same for medial
and lateral use.
In the
case of distal femoral blocks make sure that
the correct
blocks
are
selected
for
the
desired
position,
because
there
are three different blocks matching
the contours.
When using distal and posterior blocks, it is
recommended to screw tightly the posterior block first
and then the distal block. This is particularly advisable by
use of a 15 mm distal block.
The stem is additionally secured with the stem
retaining screw. The retaining screw must be tightened
with the screwdriver with torque. The required torque
is reached when the line reaches the Stem/Block position
( 4.5 Nm ).
Important
The stem retaining screw for retaining the stem is
packaged with the femoral component.
Assembling the femoral blocks
Summary of Contents for TC-PLUS Revision
Page 2: ......
Page 16: ...14...
Page 58: ...58...
Page 73: ...73...
Page 75: ...75...
Page 79: ...79...
Page 81: ...81...
Page 83: ...83...
Page 85: ...85...
Page 87: ...87...
Page 88: ...88...
Page 95: ...95 Note...
Page 96: ...96 Note...
Page 97: ...97 Note...