36
EN
Alarm/Indicator:
Device status:
Indicator light/
audio:
Impact to therapy:
Steps to remedy:
Blockage/Canister Full Alarm
The canister is full or there is
a blockage in the system.
Status/Alarm
Indicator:
Solid Yellow
The audible alarm
will sound.
The audible alarm
can be paused
by pressing the
Audio Pause
button. When
pressed the light
above the button
will illuminate.
Device will continue
to operate, but may
not provide the
prescribed therapy.
• Check your canister. If it is
full, replace the canister (see
image on page 21).
• If you have white quick click
connectors on the tubing
between your dressing and
the canister ensure that no
connector has been capped
off and the T-connector is not
blocked (see images on pages
20 and 22).
• If you have orange quick
click connectors, check the
aeration disk on your dressing
tubing, located close to the
connectors. Ensure that your
aeration disc is not blocked
(see image on pages 20 and
22).
• If you cannot locate the
source of your blockage,
contact your healthcare
provider.