20
TECHNICAL SPECIFICATION
6.2.1
Normal operating conditions
+5°C to +40°C (41°F to 104°F) at a relative humidity between 15% to 90% RH, non-condensing
but not requiring a water vapour pressure greater than 50hPa and atmospheric pressure between
700hPa to 1060hPa
6.2.2
Shipping and storage conditions
-25°C to +5°C (-13ºF to 41ºF) with any humidity level
+5°C to +35°C (41°F to 95°F) at a relative humidity up to 90%,
+35°C to 70°C non-condensing at a water vapour pressure up to 50hPa
During all transportation it is recommended that the chair remains in its original packaging.
It is recommended that the chair is stored in a cool, dry place out of direct sunlight. If cleaning is
required after storage, please see section 7.6 ‘Cleaning’ of the user manual for instructions.
Shipping and storage conditions are also contained on Hardrock Box Label:
6.3
Standards Applied
The standards that have been applied to the device are as follows:
•
BS EN ISO 12182:2012 (Assistive products for persons with disability. General
requirements and test methods).
•
IEC 62366:2015 (Application of usability engineering to medical devices).
•
BS EN ISO 15233-1 :2016 (Medical devices — Symbols to be used with medical device).
6.4
Disposal
Please observe the local laws on recycling and respect the current laws for disposal within the
community the device is being used within. If there is any uncertainty of the below guidelines,
contact your local authorities to determine the proper method of disposal of potentially bio-
hazardous parts and accessories.