8 Function Selection mode 

In measurement mode, press the 

 button between 3 and 5 seconds, 

to display [F 0]. Select to  display the function to be changed  [F

□□

]. 

Press and hold the 

 button  for  2 seconds or longer  in function 

selection mode to return to measurement mode. 

 

 

:  Some products do not have all the functions. If no function is available or 

selected due to  configuration of other functions, [-  -  -] is displayed on the sub 

display (right). 

 

8.1  Default Function settings 

The default setting is as follows. 

If no problem is caused by this setting, keep these settings. 

 

  [F 0] Display units, switch output specifications and diagnostic 

information selection function. 

Units specification 

Pressure range 

Default setting 

"Nil" or M 

ISE20C(H) 

MPa 

ZSE20C(F) 

kPa 

ISE20C(H) 

psi 

ZSE20C(F) 

  [F 1] Setting of OUT1 

Item 

Default setting 

Output mode 

Hysteresis mode 

Reversed output 

Normal output 

Pressure setting 

ISE20C 

: 0.500 MPa 

ZSE20C 

: -50.5 kPa 

ZSE20CF 

: 50.0 kPa 

ISE20CH 

: 1.000 MPa 

Hysteresis 

ISE20C 

: 0.050 MPa 

ZSE20C 

: 5.1 kPa 

ZSE20CF 

: 5.0 kPa 

ISE20CH 

: 0.100 MPa 

Delay time 

1.5 msec 

Display colour 

OUT1 ON: Green / OUT1 OFF: Red 

  [F 2] Setting of OUT2 is the same setting as [F 1] OUT1. 

 

  Other parameter settings 

Item 

Default setting 

[F 3] Digital filter setting 

0.00 s 

[F 4] Auto-preset function 

Not used 

[F 6] Fine adjustment of display value 

0% 

[F10] Sub display setting 

std (Standard) 

[F11] Display resolution setting 

1000-split 

[F14] Zero cut-off setting 

0.0 

[F80] Power saving mode 

OFF 

[F81] Security code 

OFF 

[F90] Setting of all functions 

OFF 

[F96] Number of pressurizing errors 

[F98] Output check 

N/A (normal output) 

[F99] Reset to default settings 

OFF 

 

 

9 Other Settings 

 

Snap shot function 

The current pressure value can be stored to the switch output ON/OFF 

set point. 

 

Peak/bottom value indication 

The maximum (minimum) pressure when the power is supplied is 

detected and updated. 

 

Zero clear function 

The displayed value can be adjusted to zero if the measured pressure 

is within 

±

7%F.S (

±

3.5%F.S. for compound pressure) of the zero point 

 

Key-lock function 

The key-lock function is used to prevent errors occurring due to 

unintentional changes of the set values.  
For further details refer to the operation manual on the SMC website 

(URL: 

https://www.smcworld.com

).

 

 

10 How to Order 

Refer to the operation manual on the SMC website (URL: 

https://www.smcworld.com

) for How to order information. 

 

11 Outline Dimensions 

Refer to the operation manual or catalogue on the SMC website (URL: 

https://www.smcworld.com

) for Outline Dimensions. 

 
 
 
 
 
 
 
 
 
 

12 Maintenance 

12.1 General Maintenance 

Caution 

  Not following proper maintenance procedures could cause the 

product to malfunction and lead to equipment damage. 

  If handled improperly, compressed air can be dangerous. 

  Maintenance of pneumatic systems should be performed only by 

qualified personnel. 

  Before performing maintenance, turn off the power supply and be 

sure to cut off the supply pressure. Confirm that the air is released to 

atmosphere. 

  After installation and maintenance, apply operating pressure and 

power to the equipment and perform appropriate functional and 

leakage tests to make sure the equipment is installed correctly. 

  If any electrical connections are disturbed during maintenance, 

ensure they are reconnected correctly and safety checks are carried 

out as required to ensure continued compliance with applicable 

national regulations. 

  Do not make any modification to the product. 

  Do not disassemble the product, unless required by installation or 

maintenance instructions. 

 

 

13 Limitations of Use 

13.1 Limited warranty and Disclaimer/Compliance Requirements 

Refer to Handling Precautions for SMC Products. 

Caution 

 

SMC products are not intended for use as instruments for legal 

metrology.  

Measurement instruments that SMC manufactures or sells have not 

been qualified by type approval tests relevant to the metrology 

(measurement) laws of each country.  

Therefore, SMC products cannot be used for business or certification 

ordained by the metrology (measurement) laws of each country. 

 

14 Product disposal 

This product shall not be disposed of as municipal waste. Check your 

local regulations and guidelines to dispose of this product correctly, in 

order to reduce the impact on human health and the environment. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15 Contacts 

Refer to 

www.smcworld.com

 or 

www.smc.eu

 for your local distributor / 

importer. 

 

 

URL: 

https://www.smcworld.com

 (Global)      

https://www.smc.eu

 (Europe) 

SMC Corporation, 4-14-1, Sotokanda, Chiyoda-ku, Tokyo 101-0021, Japan  

Specifications are subject to change without prior notice from the manufacturer.  

© 2021 SMC Corporation All Rights Reserved. 