Temperature Calibration Procedure

Tools Required:
Accurate Temperature Measuring Device
Water Supply

See Figure 30 for an example of the test
configuration. 
Temperature calibration must be
performed with liquid flowing through
the Tracer flowmeter.

1. Stabilize the temperature by allowing

water to run through the circuit for a
few minutes.

2. Press ON on the Tracer Flowmeter, then 

Δ

to display the temperature.

3. Record the temperature shown.
6. Plug the numbers into the following formula:

Measuring device reading  -  Tracer Flowmeter reading  =  difference
Multiply the difference x 10.

7. Add the resulting number (positive or negative) to the Temperature

Calibration Value in the Calibration Mode. Use the Calibration Mode
to change the temperature calibration value, as shown on page 7.

8. Verify that Tracer agrees with calibration thermometer and repeat

calibration procedure if needed.

Figure 30

Flow Calibration Procedure 

(continued)

Before you begin: Purge all the air from the system by running liquid
through the test apparatus. 
For best results, take readings as close to full range as possible (at least 5
gpm for the 3/8"NPT, etc.). 

1. If the display is off, push the ON button on the Tracer flowmeter.
2. Turn valve to full open position quickly and start timer

simultaneously.

3. Record flow display on the Tracer flowmeter.
4. When the liquid reaches the selected level in the container, stop

timing and close the valve.

5. Divide the volume from the container by time 

in minutes

from the

timer to determine flow rate.

6. Plug the numbers into the following formula:

Tracer reading  -  manual reading  =  difference
Multiply the difference x 200.

7. Add the resulting number (positive or negative) to the Flow

Calibration Value in the Calibration Mode. Use the Calibration Mode
to change the flow calibration value, as shown on page 7.

8. Verify and reapt calibration if needed.

Summary of Contents for Tracer Series

Page 1: ...a cooling line that provides optimum heat transfer Water flow rate greater than the point of Turbulent Flow does not provide a proportional benefit Turbulent flow tracking allows technicians to apply mathematical cooling principles to all machines in a water system Visit the Technical Documents section of www smartflow usa com for a detailed discussion of Turbulent Flow Input the percentage of gly...

Page 2: ... used Orange Flow Analog Voltage Output Violet Temp Analog Voltage Output Green Relay Common Brown Relay Normally Open Gray Relay Normally Closed Operating Instructions When power is applied to the Tracer flowmeter the software version and the Tracer Unit Code EP will appear sequentially Then the screen will be blank Press ON to activate the display There are three modes of operation for the Trace...

Page 3: ...isplayed Setup Mode will continue unless you exit Set BTU m Input Temperature Enter Setup Mode above Press ON repeatedly until the display shows BTU m in the upper left corner See Figure 7 Press or Δ to change input temperature 40 to 220 F or 4 4 to 104 4 C Press ON when desired unit is displayed Note If you set the temperature input to less than 40 F or 4 4 C the display will show off and BTU m d...

Page 4: ...r See Figure 10 Press or Δ to set the temperature at which the Normally Open switch contact will close Press ON when desired value is displayed To turn off this set point press and hold until the display shows OFF Setup Mode will continue unless you exit Set Hi Limit Temperature Switch Point Enter Setup Mode see page 3 Press ON repeatedly until the display shows F or C in the upper right corner an...

Page 5: ...Turbulent Flow display TFLOW and calculation are adjusted automatically based on process temperature and pipe size Select Relay Action to Set Points or Turbulent Flow Relay can be set to switch on user programmed set points or on Turbulent Flow Enter Setup Mode see page 3 Press ON repeatedly until the display shows rELY see Figure 14 Press or Δ to change relay action to ALARM or TFLOW Alarm settin...

Page 6: ...tup Mode see page 3 Press ON repeatedly until the display shows AO t Analog Output Temperature in the center See Figure 17 Press or Δ to set 5 or 10 Volts as the upper end of the output range Press ON when desired value is displayed Set Flow Output Upper Limit Enter Setup Mode see page 3 Press ON repeatedly until the display shows FS F Full Scale Flow in the center See Figure 18 Press or Δ to set ...

Page 7: ... entering Calibration Mode the software version will display See Figure 21 There is no adjustment to be made EEPR This function resets the calibration values to the program defaults See Figure 22 Do not reset these values This will re program the Tracer flowmeter back to pre calibrated settings The flowmeter must be recalibrated if this is done Flow Calibration Value Increase or decrease this numb...

Page 8: ...e arrow buttons until the voltmeter reads exactly 10 Volts Press the ON button to scroll to the next function Temperature See Figure 26 Connect the Violet wire to the side of a voltmeter and the Black wire to ground connection Adjust the number on the Tracer flowmeter using the arrow buttons until the voltmeter reads exactly 10 Volts Press the ON button to scroll to the next function LCD Test Scre...

Page 9: ... thermometer and repeat calibration procedure if needed Figure 30 Flow Calibration Procedure continued Before you begin Purge all the air from the system by running liquid through the test apparatus For best results take readings as close to full range as possible at least 5 gpm for the 3 8 NPT etc 1 If the display is off push the ON button on the Tracer flowmeter 2 Turn valve to full open positio...

Page 10: ...compatible chemicals Chemical Compatibility The following is a list of chemicals that are not compatible with the UDEL Polysulfone used in the Tracer Electronic Flowmeter Contact Burger Brown Engineering for more detailed information Please contact us for further information Burger Brown Engineering Inc 4500 E 142nd Street Grandview MO 64030 Tel 816 878 6675 Fax 816 878 6683 www smartflow usa com ...

Reviews: