1.FEATURES

2. SPECIFICATIONS

3. FRONT PANEL DESCRIPTIONS

4. MEASURING PROCEDURE

5. STORING AND 

    RECALLING READINGS 

6. HOW TO SET THE LEQ TIME

7. 

8.HOW TO SET THE TIME OF 

    AUTO POWER OFF

9. CALIBRATION 

10.CONSIDERATIONS

11.BATTERY REPLACEMENT

HOW TO SET THE ALARM VALUE

TABLE  OF  CONTENTS

...... ......

...

...............

..... ..... .... ..... ..... ..... ..

..

..........................

...........

............................................

................

..........

................

2.SPECIFICATIONS

Display:14mm (0.55

) LCD with backlight

Parameters measured:  Lp, Lmax, Leq , L

N

.

Measurement range: 

Lp:        30~130dB (A)
              35~130dB (C)
              35~130 dB (F)
Leq:  30-130dB (10s, 1min, 5min, 

10min, 15min, 30min

1hour, 

8hour, 24hour)

 L

N

 :      0-100%

Resolution:          0.1 dB
Accuracy:            

±

 1dB

Frequency weighting:'A', 'C', 'F' (Flat)
Time weighting :Fast (125ms) 
                             Slow (1 second)
Microphone:1/2 inch electret condenser type
Memory:     30 groups with measuring
                     conditions
Built in calibration signal:  94dB   at 1kHz 

(sinusoidal)

Frequency range:  20 to 12,500 Hz
Alarm value set:    30-130 dB
Alarm output:        LED and optional relay 

contact (NC)

PC interface:         RS232C 
Low battery indication
Power supply:       4x1.5v AAA (UM-4 

battery)

2

 inside.

11.BATTERY REPLACEMENT

11.1  When  battery  voltage  less  than  approx.  5v,  it 

is necessary to replace the batteries.

11.2  Install  a  the  batteries    4x1.5v  AAA  (UM-4) 

correctly into the case.

11.3  If    the    instrument  is  not  to  be  used  for  any 

extended period, remove batteries.

Note:  If   AC  adapter  is  used,    batteries  should  be 

taken away.

9

7

the  period  of  measurement  time,  just  press   

  for  about  8  seconds  till  "Leq"  is  on  the 

display,  release  the  key,  then  select  the  set 
time among 10s, 1min, 5min, 10min, 15min, 
30min,  1hour,  8hour,  24hour  by  depressing                   
Up  or  Down  key.  To  quit,  just  depress  any 
key except       or       key.

The  alarm  LED  is  on  when  the  instant 
measurement value is larger than or equal to 
the alarm value set. The default setting value 
at the factory is 85dB. Users can change it to 
any  value  between  30-130dB  by  following 
steps.  Depress                for  about  5  seconds, 
release  it  after  "AL"  shows  on  display,  then 
press       key or      key to preset the alarm 
value  as  desired.  To  quit,  just  depress  any 
key except       or       key.  

8.

HOW TO SET THE TIME OF AUTO POWER OFF

The default setting for auto power off at the 
factory  is  5  minutes.  That  means  the  meter 
will auto power off 5 minutes from the time 
of last key operation. Users can change it to 
any value between 1-9 minutes by following 
steps.  Depress          for  about  10  seconds, 
release  it  after  "AU

Г

O"  shows  on  display, 

then press       or       key to preset the time as 
desired.  To  disable  the  function  of  auto 
power  off,  just  preset  the  time  to  0.  It  will 

 

 

 

 

7.HOW TO SET THE ALARM VALUE?

MENU

DEL

READ

SAVE

MENU

DEL

READ

SAVE

MENU

DEL

READ

SAVE

.....................

.............

..... ......

...

...............

.....

1

2

3

4

6

7

8

9

6

7

8

SAVE

READ