background image

11. 

Wait for the CMU to sound a melodic confirmation tone before going to the next step 

 

12.

 To programme additional component(s) repeat steps 7-10 for each new component 

 

 IF CMU TIMES OUT AFTER PROGRAMMING FIRST COMPONENT(S) REFER TO 

ADDING 

ADDITIONAL COMPONENTS

 SECTION IN 

BLUE 

BELOW 

 

13.

 After you have activated the last component, wait for CMU to sound its melodic confirmation 

tone and then move the switch (14) to the LOCK position and press and release the RESET 
button (7) on the side of the CMU 
 

14. 

Test the system. Activate each component in turn, wait for alert then press RESET button. 

System is now ready for use! 
 

ADDING ADDITIONAL COMPONENTS 
 
1. 

Ensure no components are being activated when adding additional components to the CMU 

 

2.

 Place new components in the order you would like to programme them. Bed and chair pads will 

need to be on a hard surface and switched on 
 

3. 

Put the CMU into the UNLOCK position and then press the ADD/REPLACE NEW 

COMPONENT button (4) for two dings and a melodic confirmation tone then release 
 

4. 

Press the PROGRAMME COMPONENTS button (5) once and release. Wait for melodic tone 

 

5. 

Flashing number on screen is the last number that a component is currently programmed to. 

 

6. 

Activate components as “Activate First Component” above 

 

7. 

Wait for melodic confirmation tone and then put CMU into LOCK position 

 

8. 

Press and release RESET button (7) on the CMU. Test components. 

 

 

FOR FURTHER INFORMATION REFER TO THE SMART CAREGIVER 433-CMU-40/60 
INSTRUCTION BOOK SUPPLIED WITH THE WIRLESS CMU (CENTRAL MONITORING UNIT), 
OR CALL THE FRIENDLY AND KNOWLEDGEABLE STAFF AT HEALTHSAVER ON 1300 
767888 

WWW.HEALTHSAVERALARMS.COM.AU

 

 

 