background image

17

SLS Lab Basics 12 Place Microcentrifuge User Manual

11.2.2 Cleaning PC tubes

PC materials are low in chemical resistance against alkaline solutions. Avoid using neutral detergents
with pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted 
according to the instruction in the makers catalog. Use detergent with a pH between 7 and 9.

11.2.3 Autoclaving PA, PC and PP tubes

PA begins softening at about 120ºC, PC and PP at about 130ºC. Autoclave PA tubes at 115.0ºC (0.7kg/
cm

2

)

 

for 30 minutes and PC and PP tubes at 121ºC (0.1kg/cm

2

) for 20 minutes. If a certain temperature is 

exceeded, the tubes may be deformed.
When using a sterilising chamber, please operate as follows:
(1) Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform
      into an oval shape due to gravity.
(2) Remove screw nuts and inner covers to prevent from deformation or rupture.
(3) Wait until the sterilising chamber cools down to the room temperature before the tubes are removed.

11.2.4 Condition and life expectancy of tubes

The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used,
and temperature applied and so on. When the plastic tubes are used for centrifuge of ordinary aqueous 
samples (pH between 5 and 9), their life expectancies are defined as follows.
Operated at the maximum speed:
High quality tubes (PA, PC, PP): 30-50 operations
Ordinary tubes (PA, PC, PP): around 10 operations (Using in low speed can extend the tube life)
Life expectancy of tubes also depends on the pretreatment conditions such as cleaning and sterilisation, 
lifetime can be cut down.

Notice: Do not use damaged or cracked tubes.

Formaldehyde

0

0

0

PA: Polyallomer          PC: Polycarbonate             PP: Polypropylene

Summary of Contents for D2-236

Page 1: ...eland NI sales scientific labs com Republic of Ireland ROI sales scientific labs com High Speed Mini Centrifuge Before using centrifuge please carefully read this user manual for its efficient operati...

Page 2: ...nto tubes 8 7 3 Keep the tubes balanced 8 7 4 Inspect the rotor 8 7 5 Symmetrically load centrifuge tubes in rotor 8 8 Operation 9 8 1 Normal Operation 9 8 2 RCF Operation 12 8 3 Pulse Operation 12 9...

Page 3: ...bserved could result in personal injury or possible death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result...

Page 4: ...vel table Make sure the centrifuge is horizontal before running Make sure the angle between the door and cover is greater than 70 degrees when the door is open Be careful not put your fingers or hands...

Page 5: ...onstruction in accordance with the following safety standards EN 61010 1 Construction in accordance with the following EMC standards EN 61326 1 FCC Part 15 Subpart B ICES 001 Associated EU guidelines...

Page 6: ...ment The centrifuge generates uses and can radiate radio frequency energy and if not installed and used in accordance with the user manual may cause harmful interference to radio communications Operat...

Page 7: ...d installing on the slippery surface or surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if temperature exceeds 30 C 3 Keep clear of t...

Page 8: ...Door Lock Hook Foot Operation Panel Door View Window Air Vents Power Inlet Power Switch Figure 5 1 Front view of centrifuge Figure 5 2 Rear view of centrifuge 6 Door release hole SLS Lab Basics 12 Pla...

Page 9: ...m running time is 12 minutes When speed symbol is rotating indicating the centrifuge is running Time display symbol displays the ratio of working to time setting The total time setting is divided into...

Page 10: ...nder unbalanced condition even though the allowable imbalance is not exceeded 7 4 Inspect the rotor Check the rotor for corrosion or scratches before using 7 5 Symmetrically load centrifuge tubes in r...

Page 11: ...tton shift between speed and RCF display 4 Open lock button Press the button to open the door The button is not available when the centrifuge is running 5 Start Stop button Press the button to start r...

Page 12: ...e speed value rotate the program button anti clockwise to decrease speed value Fastener Figure 8 2 The rotor installation Load the rotor to the shaft to ensure rotor is in position until it is connect...

Page 13: ...ter button to modify values Release the button after 5 seconds and the centrifuge will return to normal operation mode and run according to the new value If the set time value has been modified the op...

Page 14: ...hten the rotor lid and make sure it is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last running values The values can be reset 3 Press knob and...

Page 15: ...coat of silicon grease to the rotor when it is completely dry The rotor should be regular maintained we recommend cleaning it every 3 months to ensure tube and rotor holes keep clean and then apply a...

Page 16: ...ter the cause removed 10 2 How to open the door 10 2 1 In the case of power on CAUTION The door can be opened while the power is on and the rotor stops rotating 1 Turn on the POWER switch the door loc...

Page 17: ...tube 0 2ml PCR A05P2 15000 11700 Locking nut Fastener Lid Rotor block Spindle hole Tube Locking Nut Fastener Lid Rotor Block Spindle Hole Tube Figure 11 1 The rotor profile 11 1 2 Available rotors and...

Page 18: ...iling 15 to 30 minutes O O O Ultraviolet sterilization 200 300nm X X X 11 1 3 Notice The centrifuge rotor can separate samples with density lower than 2 0g ml If the separated samples density is over...

Page 19: ...aced sideways they may deform into an oval shape due to gravity 2 Remove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilising chamber cools down to the room...

Page 20: ...igue damage We do not guarantee this centrifuge and the rotor under the following conditions even if within the guarantee period 1 Failures caused by incorrect installation 2 Failures caused by rough...

Page 21: ...dependent channel partner to a vast range of liquid handling solutions we offer pipette servicing solutions for an extensive range of brands On Site Repairs For an agreed fee the SLS Service Centre wi...

Reviews: