7.4

 

PROGRAMMING THE MTR 

The MTR has a central operating panel containing a display. The display shows two different temperatures (see 
picture 7.1) whereat the upper line states the actual temperature and the line below shoes the programmed 
temperature configured by the user. 

 

The temperature of the cooling block is configurable by pressing the                

or              

(like shown in 

picture 7.1). 

The value of the temperature is displayed in degree Celsius (°C) by factory settings, but your authorized Slee 
technician can easily change to °F. 

 

7.5

 

FIXING THE SAMPLE 

To attach the sample to the cooling block, switch on the MTR quick-freezing unit. After reaching the desired 
temperature, remove the block from the standard object bracket. Turn the block so that the screened surface of 
the block is aligned horizontally. Apply the embedding media together with the sample to the screened surface. 
When everything is solid, fix the block in the standard object clip to start cutting. 

 

7.6

 

DEFROSTING THE COOLING BLOCK 

 

 
Depending on the configured temperature and the ambient conditions (humidity), the cooling block is covered 
with ice. The MTR features a defrost function that allows the user to actively melt the ice by pressing the defrost 
button (see Figure 7.1). After the ice has melted, press the defrost button again to stop defrosting. 

The temperature for the defrost function is programmed to 5 °C (41 °F). 

 

 

 

i

 

Do not dip the cooling block into warm or hot liquids. 

Figure 7.1: Central operating panel

 

Defrost- 

Key 

Arrow  

Down 

Arrow  

Up 

Summary of Contents for MTR

Page 1: ...DESIGN MANUFACTURING MADE IN GERMANY O P E R A T I N G M A N U A L M T R BENCH TOP QUICK FREEZING UNIT ...

Page 2: ......

Page 3: ...TALLATION 6 6 3 ELECTRICAL POWER CONNECTION 6 7 OPERATION OF THE MTR______________________________________________ 7 7 1 ASSEMBLING THE COOLING BLOCK 7 7 2 MOUNTING THE COOLING BLOCK ON THE SLED MICROTOME 8 7 3 STANDARD ACCESSORIES MOUNTING MTR ON SLED MICROTOME 8 7 4 PROGRAMMING THE MTR 9 7 5 FIXING THE SAMPLE 9 7 6 DEFROSTING THE COOLING BLOCK 9 8 MAINTENANCE_____________________________________...

Page 4: ...ppen at extreme ambient humidity 2 SYMBOLS Slee bench top quick freezing units are provided with the following safety features MTR Self optimized controller for temperature accuracy Manual defrost The equipment that the appliance is equipped with and the persons who work with the appliance are waiting or repairing are responsible for a safe use Dangers warnings and cautions are marked by this symb...

Page 5: ...e 10 to 35 C Operating humidity max rel 80 non condensing Storage temperature range 5 to 55 C Storage humidity max rel 80 non condensing Dimensions width x depth x height 410 x 485 x 310 mm Hose Length 2 000 mm with connected cooling block Weight without accessories 28 kg Cooling capacity max 45 C 49 F Defrost Temperature 5 C 41 F Power Consumption 230 V 50 Hz device 300 VA 1 5 A Power Consumption...

Page 6: ...t should be positioned onto a plane vibration free surface A free ventilation of fresh air behind the device should be guaranteed The vicinity of the work area must be free of oil and chemical vapors 6 3 ELECTRICAL POWER CONNECTION Make sure that electric power is constant Connect the unit to the main supply and switch on with mains switch at the rear of the unit This should be examined during ins...

Page 7: ...to ensure a permanently secure connection the knurled screw should be hand tightened with a 17mm open ended spanner The samples can now be frozen on the front of the structured surface The supplied clamping block must be clamped in the upper position to ensure the central alignment of the cooling head Your device is now ready for use and should look as shown above i Assure that the hose can move u...

Page 8: ... as shown above 7 3 STANDARD ACCESSORIES MOUNTING MTR ON SLED MICROTOME Description Item no Pillar 37185030 Intermediate plate Cuttec S Blade holder 5 mm 45170001 Intermediate plate Cuttec S Blade holder 10 mm 45170002 Intermediate plate Cuttec S Blade holder 30 mm 45170005 Cylinder screw ISO 4762 M8 x 35 A2 70010835 Cylinder screw ISO 4762 M8 x 40 A2 70010840 Cylinder screw ISO 4762 M8 x 45 A2 70...

Page 9: ...ching the desired temperature remove the block from the standard object bracket Turn the block so that the screened surface of the block is aligned horizontally Apply the embedding media together with the sample to the screened surface When everything is solid fix the block in the standard object clip to start cutting 7 6 DEFROSTING THE COOLING BLOCK Depending on the configured temperature and the...

Page 10: ...boratory Practices 9 SERVICE Internal components should only be serviced by technicians authorized by SLEE medical GmbH If technical service or spare parts are necessary please contact your local SLEE medical GmbH distributor Please have the following information available Complete contact details Type of device and serial number Place of device and name of user Purpose of service call Delivery da...

Page 11: ...chnical specifications and or agreed char acteristics warranted SLEE medical GmbH guarantees that the device is manufactured under an ISO 9001 2015 and ISO 13485 2016 quality management system Unauthorized modification or repair by third party persons will void the warranty Only original SLEE medical GmbH spare parts must be used Guarantee claims can be put forward only if the device is used accor...

Page 12: ...SLEE medical GmbH Am Neuberg 14 55268 Nieder Olm Germany www slee de T 49 0 6136 76997 0 E mail slee de www slee de LinkedIn YouTube Instagram INS1100EN Manual_MTR_2021 10 011 00_EN ...

Reviews: