9. Maintenance

9.1 Storage & cleaning

It should be stored at room temperature. The sensor lens is the most 
delicate part of the thermometer. The lens should be kept clean at all 
times, care should be taken when cleaning the lens using only a soft 
cloth or cotton swab with water or medical alcohol, allowing the lens 
to fully dry before using the thermometer. Do not submerge any part 
of the thermometer.

9.2 LCD error messages

The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Low’ is displayed when the temperature being measured is 
outside of the settings of HAL and LAL.

‘Er2’ is displayed when the thermometer is exposed to rapid changes 
in the ambient temperature. 

‘Er3’ is displayed when the ambient temperature exceeds 
0 ºC 

(32 °F)

 or +50 ºC 

(122 °F).

 The thermometer should be  

allowed plenty of time (minimum 30 minutes) to stabilize to the 
working/room temperature.

Error 5~9, for all other error messages it is necessary to reset the 
thermometer. To reset it, turn the instrument off, remove the battery 
and wait for a minimum of one minute, reinsert the battery and turn 
on. If the error message remains please contact SKF.

‘Hi’ or ’Lo’ is displayed when the temperature being measured is 
outside of the measurement range.

9.3 Batteries

The thermometer incorporates visual low battery indication as 
follows:
•   ‘Battery OK’:  

measurements are possible.

•   ‘Battery Low’:  

 battery needs to be replaced, 
measurements are still possible.

•   ‘Battery Exhausted’:  measurements are not possible.
When the ‘Low Battery’ icon indicates the battery is low, the battery 
should be replaced immediately with AAA, 1.5V batteries. 

Please note:

 It is important to turn the instrument off before 

replacing the battery otherwise the thermometer may malfunction. 
Dispose of used battery promptly and keep away from children.  
If the device is not to be used for a long time, turn the power off, 
remove and store the batteries in a cool, dry place.

Summary of Contents for TKTL 21

Page 1: ...Instructions for use SKF TKTL 21...

Page 2: ...5 4 Technical data 6 5 Front panel and button description 8 6 Display overview 10 7 Menu overview 11 8 Measurement settings 12 9 Maintenance 13 9 1 Storage cleaning 13 9 2 LCD error messages 13 9 3 B...

Page 3: ...repair work should be taken care of by an SKF repair shop Here is a table of hazardous substance name and concentration for TKTL 21 Part name Hazardous Substances Lead Mercury Cadmium Hexavalent Chro...

Page 4: ...of the following Directive s EMC DIRECTIVE 2014 30 EU and are in conformity with the following standards EN 61326 1 2013 EN 61326 2 2 2013 Emission CISPR 11 2009 A1 2010 Class B Immunity IEC 61000 4 2...

Page 5: ...the measure target with Lens 7 and press Meas key 6 to display the surface temperature 2 The Distance Spot is 30 1 The eight laser points are the reference for the target spot size Please make sure th...

Page 6: ...me 1 000 msec 90 Optics Sense emitted reflected and transmitted energy Display Color LCD User selectable backlit display No permanently ON Displayed resolution 0 1 C F from 83 2 999 9 otherwise 1 C F...

Page 7: ...kaline type IEC LR03 Battery lifetime Min 30 hours continuous use without laser Min 3 hours continuous use with laser and white LED Switch off IR mode automatic after 60 seconds after trigger is relea...

Page 8: ...8 EN 5 Front panel and button description 6 5 1 2 3 4 Item Description 1 LCD 2 Mode key 3 Down key 4 Up key 5 Thermocouple socket 6 Meas key...

Page 9: ...9 EN 7 9 10 8 Item Description 7 Infrared lens 8 LED 9 Multi beams 10 Battery cover...

Page 10: ...10 EN 6 Display overview Default screen Multi Beams spicify the apprximate measurement area for better targetting...

Page 11: ...3 to change the High Alarm HAL or Lo Alarm LAL then press Meas key 6 to confirm it When the reading is outside the High Alarm HAL or Lo Alarm LAL limit The High or Low icon will flash and you will hea...

Page 12: ...sferred In MAX MIN mode Hold on the Meas key 6 The Bar display indicates the measuring temperature The bar shows RED color when the reading is close to maximum value and shows BLUE when close to minim...

Page 13: ...the working room temperature Error 5 9 for all other error messages it is necessary to reset the thermometer To reset it turn the instrument off remove the battery and wait for a minimum of one minut...

Page 14: ...ot aim the thermometer outside the area of interest then scan across with an up and down motion until you locate hot spot Reminders Do not use the unit to measure shiny or polished metal surfaces stai...

Page 15: ...80 to 0 95 Water 0 92 to 0 96 Lacquer matt 0 97 Ice 0 96 to 0 98 Rubber black 0 94 Snow 0 83 Plastic 0 85 to 0 95 Glass 0 90 to 0 95 Timber 0 90 Ceramic 0 90 to 0 94 Paper 0 70 to 0 94 Marble 0 94 Chr...

Page 16: ...ight of the publisher and may not be reproduced even extracts unless prior written permission is granted Every care has been taken to ensure the accuracy of the information contained in this publicati...

Reviews: