10
SKF THGD 100
5.2 Display mode
A
D
B
C
The actual pressure is indicated in the display mode. The measured value is
displayed in the corresponding unit. The MAX or FS value is indicated in the
lower part of the display.
Display Description
A. Bar graph:
Graphic indication of the actual pressure.
A pressure peak is indicated by means of a pixel (graduation mark).
B. Actual pressure reading.
The indicated value is refreshed three times per second.
C. MAX/FS:
Indicates the maximum or full scale value according to setting.
The MAX value is refreshed at three times per second.
The FS shows the upper limit of the scale (i.e 100MPa or 14 500 psi).
If OFL is displayed, this indicates that the applied pressure is outside given full
scale range.
If the OFL remains displayed, while the THGD 100 is not under
pressure, contact your local SKF Service Centre.
D. Battery life remaining
The battery life indicator shows the battery life remaining. If the battery
life indicator lashes then the batteries are nearly exhausted and should be
replaced.