SKC UMEX-100 Operating Instructions Download Page 1

Operating Instructions

 863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 • www.skcinc.com

 

 Passive Sampler for Formaldehyde

Cat. No. 500-100

Performance Pro

fi

 le

Sampling Rate:

28.6 ml/min with a relative standard deviation of 7.6% at a wind velocity of 
5 to 100 cm/sec for 15 min to 24 hrs
20.4 ml/min wind velocities < 5 cm/sec for 1 to 7 days

See the Passive Sampling Guide at www.skcinc.com for UMEx

 

100 sampling rates for other aldehydes.

Validation Range: 

0.06 to 3.0 ppm

Detection Principle: 

Formation of stable DNPH-hydrazone in the presence of formaldehyde*

Analysis Method: 

Solvent extraction and analysis by HPLC (high-performance liquid chromatography) with UV detection 

Lower Detection Limits: 

15 min: 200 ppb (0.24 mg/m

3

)

8 hr: 5 ppb (0.006 mg/m

3

)

24 hr: 2 ppb (0.002 mg/m

3

)

7 days: 0.2 ppb (0.0002 mg/m

3

)

Shelf-life:

12 mos from date of manufacture at 

 39.2 F (4 C)

Storage: 

Before use:

 

 39.2 F (4 C)           

After use:

 

 39.2 F (4 C) and analyze within 3 wks

Capacity: 

29 

μ

g/sample

Accuracy: 

± 25%, exceeds OSHA requirements

Temperature Effects: 

No e

 ect on sampling rate between 10 and 30 C

Humidity Effects: 

No e

 ect from 10 to 80% relative humidity (RH). 

Do not use sampler below 10% RH. 

Wind Velocity Effects: 

No e

 ect from 5 to 100 cm/sec

Interferences:* 

• Highly speci

fi

  c for formaldehyde. Large amounts of carbonyl compounds may reduce the uptake of formaldehyde

• Use in ozone levels < 0.5 ppm

Dimensions: 

3.4 x 1.1 x 0.35 in (8.6 x 2.8 x 0.89 cm)

Weight: 

0.38 oz (10.9 gm)

*  If sampling in an atmosphere containing formalin, see www.skcinc.com/instructions/1795.pdf for 

fi

 eld study information.

Description

The UMEx 100 Passive Sampler for Formaldehyde has a green sliding cover that opens and closes easily to start and stop sampling. Because 
of its relatively high sampling rate and sensitive analysis, the UMEx 100 can be used for 15-minute (STEL) sampling in the ppm range, 
personal monitoring up to 12 hours, and 24-hour to 7-day monitoring of contaminants found in indoor environments.

Sampling

• UMEx samplers are designed for single use only. Do NOT reuse UMEx samplers.
•  Do not use Sharpie

®

 markers or equivalent to mark label or pouch as these can off-gas VOCs.

2

1.   Open the pouch and remove the sampler. 

Do not discard the 

pouch as it is used to protect the sampler during shipment 
to the laboratory.

 

Store the pouch away from potential 

formaldehyde sources.

 

Note:

 Inside the pouch is a small length 

of tape similar to the tape inside the sampler. This tape protects 
the background of the sampler should any formaldehyde in the 
atmosphere enter the pouch. Leave the tape inside the pouch at all 
times and send along with the sample to the laboratory.

2.  Write the date, location, and 

sample start time on the label on 
the back of the sampler.

3.  Position the sampler on a 

worker’s collar for personal 
sampling or in an appropriate 
location for area sampling.

4.  Slide the sampler cover to the 

“on” position (di

 usion plate 

exposed) to begin sampling.

5.  After sampling for the desired 

time, slide the sampler cover 
to the “o

 ” position (di

 usion 

plate covered) to stop sampling.

6.  Enter the sample stop time on 

the label on the back of the 
sampler. 

7.  Immediately place the sampler 

in the original pouch and 
seal the pouch. Send it, along 
with pertinent information, 
to an accredited laboratory 
for analysis. 

Use expedited 

shipping.

4

5

7

6

&

500-100