background image

www.skcltd.com

Air Sampling Pump

Operating Instructions

This manual covers the following models:

228-9615 & 228-9620

Summary of Contents for BioLite Plus 228-9615

Page 1: ...www skcltd com BioLite Air Sampling Pump Operating Instructions This manual covers the following models 228 9615 228 9620 ...

Page 2: ...f purchase and then at yearly intervals on this date However it is the responsibility of the user to perform a risk assessment to determine the necessary frequency of servicing that is required Service Date Service Date Service Date 1 5 9 2 6 10 3 7 11 4 8 12 Please note that SKC Ltd are the only authorised service centre in the UK guaranteeing you access to the full range of genuine SKC replaceme...

Page 3: ...cltd com 007 01 013 Issue B Page 1 Contents Specifications 2 General Information 3 The BioLite Pump 4 Operating the BioLite Pump 5 Troubleshooting 6 Replacement Parts and Accessories 7 Warranty Information 8 ...

Page 4: ...el 228 9620 7 25kg 15 95 lb Enclosure rating IP20 Operating temperature 0 C to 40 C Operating humidity 0 to 95 Rh Operating noise level 61 to 66 dBA 1metre dependent on flow rate and back pressure Electrical 0 100 200 300 400 500 600 700 800 900 1000 0 20 40 60 Back Pressure mbar suction Flow Rate litre min Model Mains Supply Maximum Input Current 228 9615 120V 60Hz 3 15A 228 9620 230V 50Hz 1 25A ...

Page 5: ...pump Potential electric shock hazard A small quantity of carbon dust will be exhausted from the pump outlet in normal operation due to normal wear of the graphite pump rotor vanes The BioLite pump case is IP20 rated it is not rated as water or splashproof and therefore must not be used where it is possible for water to enter the pump casing Warning Failure to follow these guidelines will void the ...

Page 6: ... Issue B www skcltd com Carry handle Accessory mounting points Pump inlet Flow adjuster Back pressure gauge Pump outlet Fuse holder Mains inlet On off switch Cooling fan inlet Cooling air outlet grille The Biolite Pump ...

Page 7: ...ny flow calibration Switch the pump off using the on off switch at the rear of the pump 3 Connect the sample train to the pump inlet hosetail Switch on the pump using the on off switch Adjust the flow rate if required using the flow adjuster Turn the flow adjuster clockwise to reduce the flow rate and anti clockwise to increase the flow rate Allow the pump to run for five minutes for the flow rate...

Page 8: ...p and replace with an identical fuse if required Pump cannot acheive required flow rate Check for leakages in the sample train Pump vanes damaged or excessively worn Return pump to SKC Limited for service repair Pump stops due to over temperature Check for flow blockages in the sample train and rectify if required Wait for pump to cool before restarting Check that the flow adjust control is not fu...

Page 9: ...ioLite BioLite pump Replacement Parts P228920 Inlet fuse 2A for 228 9620 model only P228921 Inlet fuse 4A for 228 9615 model only P228922 Pump rotor vanes set of four If the required item is not listed contact your supplier or SKC sales on 44 0 1258 480188 SKC provide an extensive range of sampling media including filters sorbent tubes and impingers A full selection can be found in the current SKC...

Page 10: ...articular purpose and neither assumes nor authorises any person to assume for it any liability in connection with the sale of these instruments No description of the goods being sold has been made a part of the basis of the bargain or has created or amounted to an express warranty that the goods will conform to any such description The Buyer shall not be entitled to recover from SKC any consequent...

Page 11: ...a To obtain performance of any obligation under this warranty the buyer shall return the instrument freight prepaid to SKC at the following address SKC Limited 11 Sunrise Park Higher Shaftesbury Road Blandford Forum Dorset DT11 8ST t 44 0 1258 480188 f 44 0 1258 480184 6 b To obtain further information on the warranty performance contact SKC 7 This warranty is provided under English law 8 No other...

Page 12: ...obal group of companies SKC Limited 11 Sunrise Park Higher Shaftesbury Road Blandford Forum Dorset DT11 8ST United Kingdom T 44 0 1258 480188 F 44 0 1258 480184 E info skcltd com W www skcltd com 007 01 013 Issue B July 2018 ...