Hydrogen Cyanide Passive Sampler

Cat. No. 590-400

Operating Instructions

 863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 • www.skcinc.com

Introduction

The SKC Hydrogen Cyanide (HCN) Passive Sampler, speci

fi

 ed in OSHA Method 

1015, provides accurate sampling of hydrogen cyanide from 0.44 to 20 ppm for 
either 15-minute or 8-hour samples. The uptake rate for the HCN Passive Sampler 
is 28.4 ml/min. The sampler design allows the user to load 600 mg of soda lime 
sorbent (supplied in a vial) into the sampler housing 

in the 

fi

 eld

 immediately before 

sampling and transfer the sorbent from sampler to vial 

in the 

fi

 eld

 immediately 

after sampling as speci

fi

 ed in the method. Samples are desorbed with water and 

analyzed by ion chromatography/electrochemical detector. The HCN Passive 
Sampler is designed for single use.

Performance Pro

fi

 le

Sorbent/Amount:

 

Soda Lime, 600 mg (supplied in glass vial)

Housing Material:

 Nylon

Diameter:

 

1.4 in (3.5 cm)

Length 
(including clip):

 

2.5 in (6.3 cm)

Measuring Range:

 

0.44 to 20 ppm

Analysis:

 

Solvent extraction; ion chromatography (IC) with electro-
chemical detector (ELCM)

Sample Storage:

 

3 weeks at 68 F (20 C) in storage studies 

 Note:

 OSHA Method 1015 recommends that samples be sent 

to a laboratory as soon as possible. If delay is unavoidable, store 
samples in a refrigerator as a precaution.

Sample Time:

 

Validated for 15 min and 8 hrs

Sampling Rate:

 

28.4 ml/min (4.40% RSD)

www.skcinc.com

Collar Clip

Diffusion Barrier

Collar 

Clip

Vial 

Adapter

Front View 
of Housing

Back View 

of Housing

Plug

Cover

Vial containing 

soda lime sorbent 

(sorbent is white in 

appearance)

 

Summary of Contents for 590-400

Page 1: ...sorbed with water and analyzed by ion chromatography electrochemical detector The HCN Passive Sampler is designed for single use Performance Profile Sorbent Amount Soda Lime 600 mg supplied in glass vial Housing Material Nylon Diameter 1 4 in 3 5 cm Length including clip 2 5 in 6 3 cm Measuring Range 0 44 to 20 ppm Analysis Solvent extraction ion chromatography IC with electro chemical detector EL...

Page 2: ... vial adapter until secure see inset 3A Rotate entire assembly until sorbent is draining from vial to sampler Figure 3 Tap vial lightly before unscrewing and removing it from the adapter Immediately replace plug on vial adapter Store sorbent vial and cap in aluminized pouch Record start time on the sampler label Remove cover from face of sampler housing side with holes immediately before sampling ...

Page 3: ...Ensure sorbent vials are capped tightly HCN Passive Samplers are designed for single use only Do NOT reuse HCN samplers Sample Storage 3 weeks at 68 F 20 C in a clean environment in storage studies Note OSHA Method 1015 recommends that samples be sent to a laboratory as soon as possible If delay is unavoidable store samples in a refrigerator as a precaution Analysis Preparing the Sample 1 Remove s...

Page 4: ...on Cat No Hydrogen Cyanide Passive Samplers include empty sampler housings and sealed glass vials of 600 mg soda lime sorbent pk 5 590 400 HCN Passive Samplers are designed for single use only Do NOT reuse HCN samplers SKC Limited Warranty and Return Policy SKC products are subject to the SKC Limited Warranty and Return Policy which provides SKC s sole liability and the buyer s exclusive remedy To...

Reviews: