SKC 224-27 Manual Download Page 2

SKC All-in-One Low Flow Adapter/Single 
Tube Holder

Cat. No. 224-27

Introduction

The SKC All-in-One replaces the SKC Single Adjustable Low 
Flow Tube Holder and CPC for single-tube sampling with AirChek 
Series and AirLite sample pumps. All required single-tube low fl ow 
sampling accessories are included in one compact holder. The 
built-in CPC maintains a constant pressure across the holder’s 
needle valve for fl ow stability.

Operating Instructions

Form 38035 Rev 1803

Operation

See Notes before calibrating and sampling.

1.   Set pump fl ow rate to 1.5 L/min. 

See pump operating 

instructions

.

2.   Attach Tygon tubing to pump inlet.

3.   Break tips off the sorbent tube to be used for calibration and 

insert into rubber sleeve on holder (arrow on tube pointing 
toward pump).

4.   Use a calibrator to calibrate (set) fl ow rate through the 

sorbent tube. Use a small fl athead screwdriver to turn fl ow 
adjust screw on holder (counterclockwise to increase fl ow, 
clockwise to decrease fl ow). 

See pump operating instructions 

for details.

5.  Remove tube used for calibration and repeat Step 3 using a 

clean, labeled tube for sampling.

6.   Thread protective tube cover onto holder’s threaded fi tting.

7.   Use cover’s collar clip to attach holder on clothing in the 

breathing zone.

8.   Sample for the desired time.

9.   When sampling is complete, remove sample sorbent tube 

and cap. Insert sorbent tube used for calibration into sleeve 
on holder. Use calibrator to verify fl ow rate has remained 
within ± 5% of the set fl ow.

Notes

  Do not shut off fl ow completely with fl ow adjust screw or use an 

oversized screwdriver to adjust fl ow; valve or thread seat damage 
may result.

  Ensure sorbent tube fits snugly in rubber sleeve prior to 

sampling to avoid any air leakage. Two sleeves, each a different 
inner diameter (ID), are supplied.

  Calibrate before each sampling operation using the tube holder 

and pump to be used for sampling.

  If the pump fl ow faults shortly after the holder tubing is attached 

in Step 2, check that the needle valve is open by using a small 
fl athead screwdriver to turn the fl ow adjust screw on the holder 

slightly

 counterclockwise. If fl ow fault continues, check that the 

two small holes on the bottom of the built-in CPC are not blocked. 

See related bullet below

.

  Two small holes on the bottom of the built-in CPC can become 

blocked. Periodically inspect. Clean, if needed, with a small pick 
and blow particles away with a puff of air.

  For more information, see SKC Sample Setup Guides at 
  www.skcinc.com or email [email protected].

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive 
remedy. To view the complete SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.

All-in-One Low Flow Adapter/Single Tube Holder with built-in 

CPC connected to AirChek TOUCH sample pump

Tube holder with 
rubber sleeve

Threaded 

fi

 tting 

(

fi ts all SKC protective 

tube covers

)

Flow adjust screw 
(

needle valve

)

Built-in CPC 
(

for fl ow stability

)

Outlet/Tygon  tubing  to  pump       

Larger ID sleeve for 
larger OD tubes

Pump 
inlet