skcinc.com
Form 37171 Rev 200630
Page 5
Reading Charge Status on Cradle LED
LED Action
Charge Status
Red
steady
Charge in progress
Red
3 sec
Green
1 sec
(Pattern
repeats)
Approximately 75% charged
Green
steady
Charge completed/trickle charge
Notes and Cautions
•
Power off pump before removing battery.
•
Use only the SKC charging cradle Cat. No. 220-800 or 220-900 for pump.
•
Failure to follow warnings, notes, and cautions may cause injuries and voids any warranty.
•
WARNING: Substitution of components may impair intrinsic safety. AVERTISSEMENT: La substitution de
composants peut compromettre la Sécurité Intrinsèque.
•
CAUTION: The battery used in this device may present a risk of fire or explosion when heated above 212 F
(100 C) or incinerated. Replace battery with SKC Battery Pack model P75718 only. Use of another battery may
present a risk of fire or explosion.
•
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed [removed and
replaced] in an area known to be non-
hazardous. AVERTISSEMENT: Afin de prévenir l’inflammation
d’atmosphères dangereuses, ne changer les batteries que dans des emplacements désignés non dangereux.
•
Maximum charge input voltage is U
m
= 12 V
•
CAUTION: Risk of Fire and Burns. Do Not Disassemble, heat above 212 F (100 C), or incinerate. Keep battery
out of reach of children and in original package until ready to use. Dispose of used batteries promptly according
to [all state and] local recycling or waste regulations.
•
User may replace external components such as the inlet filter, battery, protective screen cover, and/or belt clip.
Service must be done by SKC to maintain performance and IS rating. Warranty is void if pumping compartment
is opened by user.
For more information on SKC pump lithium-ion (Li-Ion) battery packs, visit
skcinc.com/catalog/pdf/instructions/1918.pdf.
Turning Pump Power On/Off
Turn on:
Press the recessed power on/off button on the side of the pump (
Figure 1
). The screen will light
up and the Flow screen will be displayed.
See Navigating Menus and Screens.
Turn off:
Press the recessed power on/off button on the side of the pump.
Note: To conserve battery
power, a non-running pump will power off automatically after 5 minutes of inactivity. Also see Auto-Dim
feature/setting in Modifying Device Settings, Changing Security (Lock Out) and Auto-Dim.
Note: The power on/off button also locks/dims and unlocks/undims the touch screen during sampling.
(See Options on pump screen during sample run on page 26.)