Sizewise Crosscare Quick Reference Manual Download Page 4

Displaying Patient Weight

  —  Press 

WEIGHT/CLEAR

 button to display 

patient weight. Weight automatically 

disappears after one minute.

  —  Press 

KG/LB

 button to switch between units.

Zeroing Scale

Note: 

When zeroing scale, bed must have, mattress, 

sheets, pillows, and all accessories—and be positioned 

about 8” above lowest horizontal position. 

Patient 

should not be on bed when zeroing scale.
  1.  Ensure nothing is touching bed.
  2.  Press and hold 

ZERO/T

 button

 

until weight 

value starts to flash, then release button. 

  3.  To confirm scale is zeroed, press 

ZERO/T

 

button

 

again. “0” shows on display.

  4.  Place patient on bed.

Hold/Freeze Mode

Note: 

Hold mode allows accessories to be added or 

removed while patient is in bed.

 

1.       Wait  until  scales  are  stabilized.   

STABILIZED SCALES

 icon will display. 

 

2.   Press 

HOLD/FREEZE

 until icon appears on the 

display.

  3.  Add or remove required accessories.
 

4.   Press 

HOLD/FREEZE

 when finished to resume 

normal function.

  5.  To deactivate without fixing weight, press 

WEIGHT/CLEAR

 button.

kg

lb

Hold 

icon

Weight

Memory 

record 

indicator

Status Indicator

Stabilized 

scales icon

Zero

Hold/Freeze

Weight/

clear

Units of weight switch (kg/lb)

iBoard: Scales

iBoard Variant

  —   Some  Crosscare beds feature an iBoard with integrated mattress 

controls. Rental mattresses do not support this function.

kg
lb

2700

8604

, Rev

. 1.0

Page 4

sizewise.com    800-814-9389

Summary of Contents for Crosscare

Page 1: ...it sizewise com qr crosscare Mobi Lift handle See page 7 Foot control bed height See page 3 Foot control lateral turn See page 3 CPR lever See page 8 Foot siderail controls See page 3 iBoard See pages...

Page 2: ...LED CPR Bed height Backrest Go Button Thighrest Supervisor Panel Using the Go Button To activate the Supervisor Panel first press GO then select function Pressing GO activates functionality for 10 mi...

Page 3: ...adjustment with siderails down for a period of three minutes Lateral turn may be adjusted up to 15 degrees Press MODE and MOBILIZATION buttons simultaneously to engage siderail override feature After...

Page 4: ...lay 4 Place patient on bed Hold Freeze Mode Note Holdmodeallowsaccessoriestobeaddedor removedwhilepatientisinbed 1 Wait until scales are stabilized STABILIZED SCALES icon will display 2 Press HOLD FRE...

Page 5: ...activate after 15 minutes To extend pause period press PAUSE button again Press ON button to terminate pause period 3 To set alarm volume level press VOLUME button iBoard Bed Exit Alarm iBoard Caregiv...

Page 6: ...an be manually applied by stepping down on the red pedal Brakes automatically activate after 60 seconds if bed is plugged in Unrestricted movement Casters are unbraked Crosscare by LINET Rental Equipm...

Page 7: ...e fitting as far as it will go Bed Height Adjustment Press up down to adjust bed height ForchargingMobilePhonesandTablets Two USB ports available under the head end of bed left and right Do not plug a...

Page 8: ...ent access 3 For air Support Surfaces press CPR Engage button or turn CPR dial on blower and or open CPR valve at head of mattress if available 4 Activate CPR function a Pull CPR LEVER for backrest re...