4 Installation
4.4 Initial start-up – procedure for proper assembly
63 18 351 D3576
D3576.201.01.05.02 11.2018
31
4.4.3
Install remote interlock – optional
Explanation
The interlock is a safety device that stops laser radiation whenever the
door of the treatment room is opened. The interlock circuit must be
connected to a switch that is located near the door of the treatment
room in order to ensure automatic interruption of the laser emission.
NOTE
The installation must be performed by a qualified electrician who is
also responsible for the installation and maintenance of the electrical
system to which the SIROLaser Xtend is connected.
NOTE
Additional or different safety precautions required by the applicable
national or local regulations for the protection of dentists, assistant
personnel, or patients must also be observed.
Installation of an interlock with door switch
1.
Prepare the interlock plug by connecting the interlock cable with the
interlock plug and by opening the bridge. Please find the technical
data sheet with circuit diagram for the installation of the interlock
circuit in "Appendix C" [→ 77].
2.
Mount the prepared interlock plug into the interlock socket on the
backside of the SIROLaser Xtend.