Operation
Sirona Dental Systems GmbH
Main home screen
Operating Instructions SIROLaser Advance
5.5.4
Set-up
5.5.4.1
Activation device
In case you have purchased your SIROLaser Advance with the optional
wireless foot control you have the choice to use either the finger switch or
the wireless foot control. Select one and confirm by pressing 'OK'.
5.5.4.2
History file
After having finished a treatment all parameters will be saved and
documented in the history file - i.e. user name, application, date & time as
well as power, laser activation time, energy, and average power of this
treatment.
➢ To download the history file please insert an USB mass storage
(class version 2.0) and press the 'save' button.
Proceed as instructed on the screen.
5.5.4.3
Calibration check
The following section describes the procedure to check the calibration of
the SIROLaser Advance.
We recommend performing this check at least once a week.
Up to 7 W
In order to enable exact inspection of the performance and flawless
functioning of your SIROLaser Advance, we recommend performing
calibration check at four different power levels:
● 1 W
● 3 W
NOTICE
The finger switch is pre-set.
To be able to use the wireless foot control please see chapter "Install
wireless foot control – optional" [ ➙ 31] for further instructions.
NOTICE
A maximum of 50 treatments is possible to be stored. If the maximum is
reached the 51st treatment will replace the 1st treatment.
When downloading the history file the data will be deleted from the
control unit.
CAUTION
Use only USB Class 2.0 memory sticks.
WARNING
You must wear the supplied laser protective goggles during the entire
calibration check.