5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
5.10.1
Sanitizing with the supply module
Sanitizing must be performed:
● Regularly every 4 weeks when operating with the public drinking
water supply or if stand-alone water supply is used more than 28 days
in exceptional cases
If the simulation unit features a disinfection system, the display
Days
until next sanitation
appears on the EasyPad or touchscreen. It first
appears three days before the date of sanitation.
● After longer periods of disuse (> one week)
● If the germination index exceeds 100 germs per milliliter, see
“Microbiological water test”.
● After changing from stand-alone water supply to public drinking water
supply and DENTOSEPT P disinfectant
The sanitizing process takes at least 24 hours.
Additional devices, sanitation with disinfectant
Preparation
The following preparations should be made prior to beginning sanitation.
1. Rinse the water lines. Activate the Sprayvit for at least one minute.
2. Set all burr drives and the SIROSONIC L scaler to the maximum
water flow rate. The water regulators under the instrument holder
must be turned counterclockwise.
3. Put all instruments and suction hoses in place.
41d
B
10:45:03
San
3d
NOTICE
Additional devices connected to the external device connection must
not be sanitized with the simulation unit.
Additional devices can be damaged. Residues of DENTOSEPT P
disinfectant can remain in the additional devices.
➢ Disconnect any additional devices from the simulation unit prior to
sanitation.
IMPORTANT
The water flow rate to the instruments is not checked by the simulation
unit.
➢ Make sure that the maximum water flow rate is set for all
instruments to be purged.