5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.3 Instruments and instrument hoses
båÖäáëÜ
5.3.5
Change the cotton wool roll on the turbine hose and oil
collector
Return air containing a small amount of highspeed handpiece oil is
emitted at the unit end of the highspeed handpiece hose. This oil is
collected in a drip container by the cotton wool roll.
TS dentist element
1. Slide the drip container down and remove the cotton wool roll.
2. Insert a new cotton wool roll and slide the container back up.
CS dentist element
Cotton wool roll on the highspeed handpiece hose
The instrument connections are located below the instrument holder. The
holder is fixed on the front edge with two internal clamps on the dentist
element.
1. The instruments must be removed from the instrument holder.
Remove all instruments one after the other and allow these to hang
down facing forward.
2. Lift the instrument holder by the front edge until the clamps are
released and the holder can be removed.
3. Slide the drip container on the highspeed handpiece hose up and
remove the cotton wool roll.
4. Insert a new cotton wool roll and slide the container back down.
5. First insert the rear edge of the instrument holder into the groove on
the dentist element. Then push the holder forward and down until it
clicks into place.
6. Place the instruments in their holders. Make sure that the instrument
hoses are located in the guide rollers of the swivel arm.