5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.3 Instruments and instrument hoses
Operating Instructions Simulation Unit SIM MODULAR
5.3.4.3
Disinfecting/sterilizing Mini L.E.D. curing light
Care, cleaning, and disinfecting agents
1. Remove the connection cable of the Mini LED by turning the
handpiece.
2. Pull out the light guide and remove the glare shield.
3. Sterilize the light guide and the glare shield at 134° C, 2 bar for 3 min.
4. Disinfect the handpiece of the Mini L.E.D.
5. Screw the sterilized light guide and glare shield back onto the Mini
L.E.D.
6. Reconnect the handpiece of the Mini L.E.D. to the connecting cable.
The following points should also be observed when operating the Mini
L.E.D.:
● Always use the glare shield to protect your eyes.
● Check the light guide after each use. Make sure that the light guide is
in perfect condition.
● There should be no traces of composite material on the light guide.
Immediately remove any residue.
● If you find any damage, replace the light guide, since damage will
impair its performance considerably.
5.3.4.4
Cleaning/disinfecting the SiroCam F / AF / AF+ intraoral camera
Due to its shape, the SiroCam AF / AF+ intraoral camera complies with
hygienic requirements and therefore has no spots that are difficult to
reach. It can be wipe disinfected.
Care, cleaning, and disinfecting agents
Intraoral camera, lens window
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 148].
134°C 274°F
3 min
SiroCam AF
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ → 148].
NOTICE
The lens window is sensitive to scratches.
Deep scratches in the lens window impair image quality.
➢ Protect the lens window against scratching. Disinfect it with a soft,
lint-free cloth.