5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.3 Instruments and instrument hoses
båÖäáëÜ
5.3.4
Lubricate, disinfect/sterilize the simulation unit
5.3.4.1
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
List of treatment instruments, E motor, without implant and HF surgery
● BL E motor
● BL ISO E motor
● Straight and contra-angle handpieces in various versions
● Highspeed handpieces
● 3-way syringeSPRAYVIT E
● Standard3-way syringe, see "Cleaning and sterilizing the standard 3-
● SIROSONIC L Scaler
5.3.4.2
Cleaning and sterilizing the standard 3-way syringe
The Standard 3-way syringe can be cleaned under running water. Spray
and wipe disinfection can be used. The jet can be removed from the tip
for sterilization.
➢ Press on the ring-shaped unlock button.
The jet adapter is released. The jet will fall out.
To attach the jet, press it in until it clicks audibly into place.
Multifunctional syringe, air jet falling out
H O
2
134°C 274°F
3 min
WARNING
The nozzle of the Standard 3-way syringe must lock into the holder.
Otherwise, the jet could fall out during treatment.
➢ Prior to treatment, ensure that the jet is securely in place.