4 Operation
Sirona Dental Systems GmbH
4.7 Stand-alone water supply
Operating Instructions Simulation Unit SIM MODULAR
4.7
Stand-alone water supply
If the water quality of the public drinking water supply is not suitable, the
simulation unit can be operated with a self-sufficient water supply.
In this case, food-grade water is mixed with DENTOSEPT P disinfectant
in a 100:1 ratio (1 liter of water, 10 ml DENTOSEPT P) and filled into the
fresh water bottle or disinfectant tank. The disinfectant will reduce the
microbial growth in the water lines.
Please note that drinking water, e.g., water from bottles, must never be
filled into the disinfectant tank (due to the minerals contained therein).
Aqua purificata or aqua destillata can be used as distilled water. Water
from bottles can be used without any restrictions for the fresh water bottle.
Regularly sanitize the water lines (every 4 weeks) and always after
switching to the public drinking water supply from the stand-alone water
supply (see “Sanitization”).
The functionality of the stand-alone water supply is only available if the
simulation unit is equipped with a fresh water bottle. The simulation unit
must be equipped with the integrated water disinfection system.
Water lines, microorganisms, stand-alone water supply
CAUTION
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one’s
health.
➢ Sirona recommends to never operate the simulation unit without
disinfectant.
➢ Mix fresh water for the stand-alone water supply every day. At the
end of the work day, the fresh water bottle must be emptied and the
remaining water must be flushed out of the disinfectant tank.
➢ Check the amount of germs in the water of the simulation unit at
regular intervals. In particular, if no disinfectant is available or if the
water has to be used without disinfectant. See “Microbiological
water test”