5 System setup
Sirona Dental Systems GmbH
Editing an exposure template
Operator’s Manual SIDEXIS XG
● Explanation
Depending on the image type involved, there are certain settings that are
useful for diagnostic image display.
The possibility of having these settings performed automatically by
SIDEXIS XG in the screen display of the corresponding image type has
been created here.
● Configuration
The following screen displays can be set depending on the image type:
-
"Filter"
Up to three filters can be specified for the screen display here.
Three list boxes are available for this purpose.
The list boxes are processed from top to bottom.
Example:
After an image view has been optimized with the
"Sharpen"
filters, the
"Reduce noise (median)"
filter is used to reduce the increased noise
level of the image view.
-
"Settings"
-
"Brightness"
/ The brightness is set here.
-
"Contrast"
/ The contrast is set here.
-
"Zoom"
/ The display size is set here. For setting
"Zoom"
: 1:1, the
zoom is adjusted to match the selected resolution.
-
"Yes"
check box
The settings are activated for the corresponding image type by
selecting the
"Yes"
check box.
"Configure"
/"Template..."
Starts the configuration workspace for exposure templates that can be used
for exams. See Section "Editing an exposure template [
"Usual X-ray tube settings (intraoral radiography only)"
The customary settings for the tube assembly used for the intraoral X-rays are
entered in the
"Tube voltage in kV"
and
"Anode current in mA"
boxes. These
values are required exclusively for documentation in the image database.
5.2
Editing an exposure template
Explanation
A special configuration workspace is used to edit existing exposure templates
and to create new ones.
Activating the function
1. From the menu bar
"[U]tilities"
, select the menu item
"Configure
[s]ystem..."
.
2. Press the
"Exposure..."
button.
ª
The
"Exposure"
dialog box appears.
3. In the
"Configure"
field, click the
"Template..."
button.
ª
The configuration workspace appears.
CAUTION
The values entered as tube assembly settings do not affect the actual
device settings in any way.