background image

Getting Started

The Schick WiFi system acquires and transmits intraoral X-ray images wirelessly over WiFi. There are two main parts of the system: the

 Schick WiFi 

Interface

, which provides power to the CDR Elite Sensor and transfers intraoral X-ray images wirelessly, and the 

Schick WiFi Docking Station

, which 

recharges the battery inside the Schick WiFi Interface and is also used during initial setup. 

Refer to the graphics below for the location of the power button, visual indicators, and connections to your Schick WiFi system. The LED at the tip 
of the Schick WiFi Interface provides continuous feedback on battery strength and diff erent fl ashing patterns to describe system status.

Installing and Operating the Schick WiFi System

password for your WiFi network. 
Press Connect and wait for the confi rmation 

 

»

message.
Click OK to confi rm the wireless network the 

 

»

Schick WiFi Interface is connecting to.
Click OK when the message, “New network 

 

»

settings were written successfully” is displayed.
Remove the Schick WiFi Interface from 

 

»

the Docking Station and close the WiFi 
Confi guration Utility. 
Disconnect the USB cable from the Docking 

 

»

Station.
Launch the Wireless Monitor by either starting 

 

»

your imaging software (which opens the 
monitor automatically), or by clicking on 
Windows Start>Programs>Startup>Wireless 
Monitor. 

(If a Windows Firewall message appears, 

click Unblock.)

Pairing

5. 

Make sure that the Schick WiFi Interface is ON 

 

»

and its LED is fl ashing. 
Click on the Wireless Monitor icon located in 

 

»

the Windows System Tray and select Schick 
WiFi Interface Management.
Click the Schick WiFi Interface to be paired. 

 

»

(It 

may take up to 45 seconds to appear on the list 
after it is turned on.)

 

Click the Pair button.

 

»

When the pairing message appears, press the 

 

»

Schick WiFi Interface button to confi rm pairing.
The Schick WiFi Interface Management 

 

»

window will show that the Schick WiFi 
Interface is paired.

Verify Installation and Operation

6. 

Prior to initial use, charge the Schick WiFi 

 

»

Interface in the Docking Station for 2 hours.
Start your particular imaging program. 

 

»

Connect the CDR Elite sensor to the Schick WiFi 

 

»

Interface.
Create or open a sample X-ray exam and place 

 

»

a radio-opaque object like a clip on the active 
side of the Sensor.
Take several sample exposures and verify 

 

»

image quality.

Install the WiFi Confi guration Utility

1. 

Only one computer is needed to confi gure the 

 

»

Schick WiFi Interface.
Insert the Schick WiFi CD into your CD-ROM or 

 

»

DVD drive. The Schick WiFi Start page should 
display automatically. If it doesn’t, browse the 
CD and double-click on Setup.exe.
Click the Install WiFi Confi guration Utility 

 

»

button.  

(If prompted, install the .NET 

Framework.)

Click Next at the Welcome screen and at the 

 

»

two screens that follow.
Click Close when installation is complete.

 

»

Install the Schick WiFi Wireless Software

2. 

Install this Schick WiFi driver on every PC used 

 

»

to acquire images.
Click the Install Wireless Software button on 

 

»

the Schick WiFi Start page.
Click Next at the Welcome screen and at the 

 

»

screen that follows.
Click Finish and close the Schick WiFi Start 

 

»

page.

Connect the Schick WiFi Docking Station

3. 

Connect the power supply cable to the Schick 

 

»

WiFi Docking Station.
Connect the power supply adapter to a wall 

 

»

outlet in the operatory.
Open the compartment on the Docking 

 

»

Station and connect the USB cable to the 
Docking Station and to the imaging PC 
with the WiFi Confi guration utility installed. 

(Please wait a few seconds while drivers install 
automatically.)

Confi gure the Schick WiFi Interface 

4. 

Insert the Schick WiFi Interface into the 

 

»

Docking Station and wait for the LED to start 
fl ashing.
Start the WiFi Confi guration Utility (Windows 

 

»

Start>Programs>Schick Technologies, 
Inc.>Schick WiFi>WiFi Confi guration Utility).
Select your wireless network from the 

 

»

Networks Available window and type in the 

Schick WiFi at a Glance

Power On / Off 

RFID Tag

Sensor and 

Docking Station 

Connection

LED

USB Connection

USB LED

Power Connection

Power LED

Schick WiFi Interface Connection LED

To install the system, please follow the steps in this Guide (full details are in the user manual). Installation includes 

confi guring

the Schick WiFi Interface to connect to your offi

  ce wireless network, and 

pairing

 it to a desktop or laptop PC. Pairing connects the 

Schick WiFi Interface to an imaging PC, enabling X-rays to be sent there wirelessly.

Schick WiFi Interface LED Color

Battery Power Level

Green

Full 

Yellow to Amber

Medium

Red

Low

Schick WiFi Interface LED Flashing

Status

Solid, purple

Powering on and booting up

Short fl ashes

Searching for WiFi connection

Long fl ash followed by two short ones

WiFi connected, waiting to pair (unpaired)

Solid, no fl ashes

Pairing with computer (“confi rming”)

Short fl ash every second

Paired, no Sensor connected

Glows from dim to bright (“breathing”)

Paired, Sensor connected

Rapid fl ashes

Transferring image

Battery Power and Status Indications