7 Program group sinus views
Sirona Dental Systems GmbH
7.4 S4 Maxillary sinuses with two views in one image (linear slice orientation)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
48
D 3352.201.01.18.02
7.4
S4 Maxillary sinuses with two views in one image (linear slice
orientation)
S4 Maxillary sinuses with two views in one
image (linear slice orientation)
(to allow for a certain depth localization).
•
Blue
bite block or contact segment.
•
Fit temporomandibular joint supports "
1
" and "
2
"
without ear holders, but with contact pads
.
•
Head inclination using the FH.
The line displayed on the touchscreen head icon in-
dicates the reflection line of the light localizer.
Do not release the exposure release button until the
message “R button, confirm exposure data” appears
in the comment line.
(Radiation is released automatically twice).
NOTE
i
For devices until October 2006:
Please observe the orientation (right/left) of the temporo-
mandibular joint supports specified on page 28.
CEPH
PAN
PAN
1260
10
68kV
8mA
?
S4
14,1
s
14,1
s
TS
Ready for exposure