13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.1.4
Ceph - Primary diaphragm
13.2.1.4.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
Adjustment 1
✔
The OP XG needle phantom is
not
in the pan bite block holder on the
unit.
✔
The OP XG ceph adjustment phantom is in the pan slot on the
✔
The ceph sensor is plugged into the ceph slot on the unit.
✔
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
"2D Adjustment"
, click on the
"Ceph -
Primary diaphragm"
element (S010.3).
The
"Ceph - Primary diaphragm"
menu is displayed in the action
area.
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (64 kV/16 mA; 6.1 s) [ → 206].
SIDEXIS XG calculates the adjustment values S1, S2, S3, and
S4 automatically from the exposure and enters them in the text
boxes of the
"Ceph - Primary diaphragm"
menu.
The exposure is displayed in the exposure window in SIDEXIS
XG.
NOTICE
Risk of damage to unit
It is essential that the OP XG needle phantom is removed from the bite
block holder on the device before a ceph exposure is taken; otherwise,
the combisensor could collide with the phantom.