63 03 452 D3352
164
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4.9
S018: Test step 10 (occlusal bite block only)
Setting the PWM values for the height adjustment creep speed
In user mode, check whether a correction is required for the speed for
upward and downward movement (creep speed). Here the bite block
must be deflected in such a way that the creep speed is effective in the
relevant direction (observe green arrows). The PWM index values are set
correctly if:
● the drive runs at approximately the same speed in both directions
● the drive spontaneously starts without jolting and concurrent to key
actuation
● the travel speed does not exceed 5 mm/s.
* Factory setting
Adjusting and saving the PWM index
values
1. Call service routine S018.10 [ → 125].
The PWM index for upward travel "U" (up) is displayed in
selection field 1.
The PWM index for downward travel "D" (down) is displayed in
selection field 2.
The display of patient symbol key 1 (G) indicates whether or not
the occlusal bite block is connected (see table).
2. Select the desired index values using the arrow keys in selection
fields 1 and 2 (see table).
3. To save the index values, first press the Memory key (the R key lights
up) and then the R key.
4. Exit the service routine [ → 129].
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) lit
Occlusal bite block
connected
not lit
Occlusal bite block not
connected
Selection field
Parameters
Range of values
1
U (upward travel)
1U – 15U*
2
D (downward travel)
1D – 15D*