11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.4.5
S018: Test step 6
Undoing the minimum travel height setting
Undoing the minimum travel height: XG 3D
1. Call service routine S018.6 [ → 125].
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
2. To undo the minimum travel height setting, touch the Memory key (B)
(R key (C) lights up) followed by the R key (C).