Sirona Dental Systems GmbH
11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.3.6
S017: Test step 8
Selecting the kV/mA level series
* Factory setting
1. Call service routine S017.8 [ → 125].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
Selection field
Code
Function
1
1A
16 mA series for pan
16 mA series for ceph
2A
8 mA series for pan
16 mA series for ceph
4A*
8 mA/16 mA series for pan
16 mA series for ceph