background image

Technical Information  

TI0387R9

 

100 HUNTER PLACE, YOUNGSVILLE, NC 27596  USA 

Ph: (919) 554-2244, (800) 356-7311 • Fax: (919) 554-2266, (800) 899-8181  

• Web: www.sirchie.com • Email: [email protected] 

 

6. In most cases, it is not necessary to adjust the sensitivity of the instrument. However, if the 

alarm sounds before a leak source can be found, it is likely the air is contaminated with 
heavy concentrations of gas. You may desensitize the instrument by turning the sensitivity 
knob counter-clockwise to a lower setting in order to pinpoint the leak location. 

NOTE: Joint compounds used on newly installed piping may contain combustible solvents and 
could cause a false alarm. 

Status and Leak Intensity Indicators 

Five (5) small LED status indicators are located on the right front of the instrument, in a vertical 
column arrangement. Six (6) Leak Intensity indicators are arranged in a horizontal row, below 
the status indicators and Sensitivity control knob. 

From top to bottom, the Status LED display functions are: 

 

Power (red):  Illuminates when the instrument is ON. 

Low Battery (red):  Illuminates when a battery charge is 
needed. The battery pack must be fully recharged before 
the instrument is used again. 

Charge (yellow):  Illuminates during the recharge process. 

Ready (green): Illuminates when the charge cycle is 
complete. A fully charged battery pack provides up to three 
(3) hours of continuous operation. 

Fault (red):  Illuminates only if there is a problem with 
charging the battery pack.  

Leak Intensity Indicators (red): Illuminate from left to right, 
indicating stronger leak intensity as more LEDs are lit. 

 

Cleaning and Maintenance 

 

Do not allow dirt or grease to obstruct the charger input jack on the side of the instrument. 

 

Use only a soft cloth dampened with mild soap to clean the body of the instrument. 

 

Do not subject the instrument to extreme mechanical shock, exposure to strong magnetic 
fields or extreme temperatures.  

 

Do not drop the battery pack or subject it to mechanical shock or extreme temperatures. 

 

Never immerse the instrument or the probe into a liquid. 

 

Figure 5: Status and Leak Intensity Indicators

 

Reviews: