Table 35: Parameter 14 values
Value range
(mg/m
3
/mA)
Description
Default
000 ... 099
Value for the decimal part of slope
𝜷𝜷
.
000
9.4.11
Parameters 15 and 16: IEU to 20 mA scaling
Parameters 15 and 16 control the maximum measurement range by setting an IEU value that
corresponds to the 20 mA output signal. This allows you to set the measurement range more accurately
than with parameter 3, which controls the measurement range by setting a multiplier in relation to the
baseline value.
The IEU scale ranges from 0 to several million units. Because of this, setting the IEU value is done with
two parameters. Parameter 15 sets the IEU value in millions, and parameter 16 in thousands.
Table 36: Parameter 15 and 16 functions
No. Parameter
Description
5.
(15) IEU units in millions
Parameter value represents × million IEU units
6.
(16) IEU units in thousands
Parameter value represents × thousand IEU unit
This means that for setting 20 mA to correspond to, for example, 125 020 000 IEU, the parameter values
would be as shown in Table 37:
Table 37: Example parameter values for 125 020 000 IEU
No. Parameter
Parameter value
5.
(15) IEU units in millions
125
6.
(16) IEU units in thousands
020
Adjusting the measurement range with parameters 15 and 16 overrides the previous
range settings done using parameter 3.
Parameter 15: IEU units in millions
Parameter 15 sets the million part of the IEU value that corresponds to 20 mA output signal. Parameter
value represents × million IEU units.
Table 38: Parameter 15 values
Value range
(million IEU)
Description
Default
000 ... 999
Sets the million part of the IEU value
004