10. After testing is finished, result appears in the meter
Display.
11. Remove test strip from meter and discard into an
appropriate waste container. Meter turns off. Result is stored
in the Memory.
Note:
-If test strip has been out of the vial too long before
testing, an error message appears. Remove and discard old
test strip. Use new test strip for testing.
-Do not place blood drop on top of test strip.
-Removing the test strip before result is displayed cancels
the test. An error message appears. Result is not stored in
Memory. Retest with a new test strip. Do not remove before
result is displayed.
-Only professional can collect venous samples.
System and Laboratory Testing
When comparing results between the product and a
laboratory system, meter blood tests should be performed
within 30 minutes of a laboratory test.
View Averages (7-, 14-, and 30-Day)
1. With meter off, press and release Set Button.
2.
Display scrolls through the 7-, 14-, and 30-day Averages.
Meter turns off after 2 minutes if Set Button is not pressed. If
there are no Average values, three dashes are displayed for
7-, 14-, and 30-Day Averages.
4.4 Meter Memory