f.

Do not work alone on high voltage circuits. Make certain that someone ca-
pable of giving aid is nearby and alert.

g.

Do not handle the Instrument, its test leads, or the circuitry while high voltage
is being applied.

h.

Hands, shoes, floor, and workbench must be dry. Avoid making measure-
ments under humid, damp, or other environmental conditions that could
effect the dielectric withstanding voltage of the test leads or Instruments.

i.

Do not change switch settings or test lead connections while the circuit is
energized. A mistake could result in damage to the Instrument and personal
injury.

j.

Locate all voltage sources and accessible current paths before making con-
nections to circuitry. High voltages may appear unexpectedly or in unex-
pected locations in faulty equipment. An open bleeder resistor, for example,
may result in a capacitor retaining a dangerous charge.

k.

Make certain that the equipment being worked upon is properly grounded
and fuses are of the proper type and rating.

l.

Double check switch positions and jack connections before applying power
to the Instrument.

m. Remain alert for low voltage circuits which may be floating at high voltage

with respect to earth ground and for composite voltages (AC + DC) such as
are found in r-f amplifiers. The floating voltage or composite voltage must not
exceed the Instruments rated maximum circuit-to-ground voltage.

n.

Do not make electrical measurements where the air may contain explosive
concentrations of gas or dust such as in mines, grain elevators, gasoline
stations or in the presence of charging batteries until determined to be safe
by qualified personnel. Note that even metallic dusts can be explosive.

o.

No General Purpose VOM is to be used to make electrical measurements on
blasting circuits or blasting caps.

4.2

Polarity Reversing

The function switch provides a convenient means to reverse polarity of the inter-
nal battery to facilitate testing semiconductor devices. The +DC and -DC posi-
tions transpose internal connections to the COMMON and +jacks and therefore
can also be used to reverse polarity of external voltages and currents connected
to these jacks (only). Because polarity reversing momentarily interrupts the cir-
cuit, it should not be done while measuring current or voltage without first turning
off the circuit power. When making measurements on the 50

μ

A/250mV, 1 volt, or

10 amp range, polarity can be corrected only by reversing the test leads.

4.3

Measurement of Unknown Voltage or Current

Sometimes the approximate voltage or current to be measured is known and the
correct range may easily be selected. It is often the case, however, particularly
when working upon faulty equipment, when the voltage or current is either un-
known or may be much higher than normal and anticipated. Whenever working
upon unfamiliar or unknown equipment, always begin a measurement using the
highest range available on the Instrument. Once it is determined that the voltage
is within the limits of a lower range, change to the lower range.

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Summary of Contents for 260 series 5M

Page 1: ...Simpson 260 Series 8P Volt Ohm Milliammeters INSTRUCTION MANUAL www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Page 2: ...rrent from that part to ground exceeds 0 5 milliampere when mea sured with an appropriate measuring instrument defined in Section 11 6 1 of ANSI C 39 5 Warranty and Returns SIMPSON ELECTRIC COMPANY wa...

Page 3: ...Measurement 16 4 5 1 DC Voltage Measurement 0 250 mV Range 16 4 5 2 DC Voltage Measurement 0 1V Range 16 4 5 3 DC Voltage Measurement 0 2 5 Through 0 250V Range 17 4 5 4 DC Voltage Measurement 0 500V...

Page 4: ...5 5 2 Battery Replacement 27 5 3 Fuse Replacement 27 5 4 Test Lead Inspection 28 5 5 Care 28 www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Page 5: ...t circuit is common to all ranges When actuated by an overload the relay contacts which are in the COMMON CIR CUIT latch open until the reset button on the front panel is depressed The reset button is...

Page 6: ...Instrument in most cases of accidental overload However no overload protection system is com pletely foolproof and misapplication on high voltage circuits can damage the Instrument Exercise care and...

Page 7: ...racy since the Instrument is calibrated in this position 1 8 Test Leads Each Instrument is furnished with one pair Catalog Number 00043 of probe type test leads one red and one black each 48 inches lo...

Page 8: ...6 to 36 dB 20 to 50 dB Reference Level With zero dB power level equal to 1 mW across a 600 line 11 BATTERIES Voltage 1 5V 9V NEDA No 13F 1604A 12 FUSE F1 1A 250V type 3AG quick acting LittlefuseType 3...

Page 9: ...ese symbols appear throughout this publication where applicable and are defined on the inside front cover of this manual under NOTES AND SAFETY INFORMATION Please adhere to these instructions to ensur...

Page 10: ...0 or 250 amperes NOTE Instructions are furnished with each Amp Clamp 2 INSTALLATION This section contains information and instructions for the installation and ship ping of the Simpson 260 8P Included...

Page 11: ...replacing the 1 5 volt D cell 2 5 Operating Position A handle is attached to the side of the Instrument case The handle may be used to support the Instrument in a convenient sloping position for easy...

Page 12: ...he COMMON jack is used in conjunction with the black test lead as the refer ence point for the measurement of all the functions with the exception of the 10 A range Refer to Section 4 for details 6 Po...

Page 13: ...ive equipment and procedures This section of the manual contains information required to use and operate the 260 8P in a safe and proper manner 4 1 Safety Precautions The following precautions are sug...

Page 14: ...in explosive concentrations of gas or dust such as in mines grain elevators gasoline stations or in the presence of charging batteries until determined to be safe by qualified personnel Note that even...

Page 15: ...d test lead into the 50 Amps 250mV 10A jack c Set the range switch at 50 Amps COMMON position with 50V d Connect the black test lead to the negative side of the circuit being measured and the red test...

Page 16: ...gures directly For the 25V range use the 0 250 figures and divide by 10 NOTE Turn off power to the circuit and wait until the meter indicates zero before disconnecting the test leads 4 5 4 DC Voltage...

Page 17: ...charged Connect the black test lead to the negative side of the circuit be ing measured and the red test lead to the positive side of the circuit e Turn on power in circuit being mea sured f Read the...

Page 18: ...d 2 5V 10V 50V or 250V When in doubt as to actual volt age present always use the highest voltage range as a pro tection to the Instrument If the voltage is within a lower range the switch may be set...

Page 19: ...The meter will not indicate if the switch is incorrectly set to a DC po sition b Set the range switch at the 250V 500V 1000V position c Plug the black test lead into the COMMON jack and the red test...

Page 20: ...d multiply by 100 4 7 Output Voltage Measurement Certain situations necessitatethe measuring of the AC component of an Output Voltage where both AC and DC voltage levels exist This occurs primarily in...

Page 21: ...ack connections c Use operating instructions for AC VOLTAGE MEASUREMENT 0 2 5 10 25 50 250V RANGES paragraph 4 6 1 Do not use the 500V or 1000V ranges for decibel readings d Read decibels on the botto...

Page 22: ...it conti nuity 4 9 1 Direct Current Measurement 0 50 A Range a Set the function switch at DC b Plug the black test lead into the COMMON jack and the red test lead into the 50 AMPS 250mV jack c Set the...

Page 23: ...OM in series with the circuit connecting the red test lead to the positive side and the black test lead to the negative side NOTE The function switch has no effect on polarity for the 10 AMPS range d...

Page 24: ...nce a Before measuring resistance in a circuit make sure the power is off to the circuit being tested and all capacitors are discharged Disconnect shunting components from the circuit before measuring...

Page 25: ...given position For example a diode which measures 80 ohms on the R X 1 range may mea sure 300 ohms on the R X 100 range The difference in values is a result of the diode characteristics and is not ind...

Page 26: ...Recommended replacement batteries are 1 5V D cell NEDA 13F Duracell MN or Eveready No 95 9V battery NEDA 1604A Duracell MN1604 or Eveready 522 c To install or replace a battery de energize and disconn...

Page 27: ...netic interferences d Verify Instrument accuracy by performing operational checks using known accurate stable sources If proper calibration equipment is not available contact the nearest Authorized Se...

Reviews: