User manual - RATEMETER STI-73
Register Write
Range
Register description
38h
Yes
0 ÷ 2
“ALArmS”
parameter in
“rELAy”
menu:
0
- no changes;
1
- on;
2
- off
39h
3Ah
Yes
Yes
0 ÷ 999999 “
SEt P2
” parameter in “
rELAy
” menu, no decimal point included
70h
Yes
0 ÷ 1
“
modE
” parameter in “
HOLd
” menu (type of detected changes):
0
- peaks;
1
- drops
71h
72h
Yes
0 ÷ 999999 “
PEA
” parameter in “
HOLd
” menu (minimum detectable change, no
decimal point included)
73h
Yes
0 ÷ 199
“
timE
” parameter in “
HOLd
” menu,
maximum peaks' (or drops')
display time expressed in seconds
74h
Yes
0 ÷ 1
“
HdiS
” parameter in “
HOLd
” menu (the type of values displayed on
the display
):
0
- current measurement value;
1
- peaks (or drops) values
75h
Yes
0 ÷ 1
“
H rEL
” parameter in “
HOLd
” menu (the control mode of relay R and
LED R)
:
0
- control depends on current measurement values;
1
- control depends on peaks (or drops) values;
D8h
Yes
0 ÷ 1
“
F or P
” parameter in
“inPUt”
menu
(kind of value visible on the
display):
0
- frequency;
1
- period
1
- it is recommended to read these registers simultaneously – in 2-registers frames. If single registers are read, data
errors are possible because of changes of read value between successive registers readings.
2 - after writing to register no 20h the device responds witch an “old” address in the message.
3
- after writing to register no 22h the device responds with the new baud rate.
4
- the value of the “mbAccE” parameter is also connected to write to this register, so it is possible to block a writes,
but impossible to unblock writes via RS-485 interface, The unblocking of the writes is possible from menu level only.
10.2.
TRANSMISSION ERRORS DESCRIPTION
If an error occurs while write or read of single register, then the device sends an error
code according to Modbus RTU specifications (example message no 5).
Error codes:
01h
- illegal function (only functions 03h, 06h and 10h are available),
02h
- illegal register address
03h
- illegal data value
08h
- no write permission ( see: “
mbAccE
” parameter)
35