Guidance and manufacture’s declaration
–
electromagnetic immunity for
all eQuiPmeNt and sYstems
Guidance and manufacture’s declaration – electromagnetic
immunity
The SPO 55 Pulse Oximeter is intended for use in the electromagnetic environment
specified specified below. The user of SPO 55 Pulse Oximeter should assure that it is
used in such an environment.
immunity
test
ieC60601 test
level
Compliance
level
electromagnetic envi-
ronment-guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 6 KV contact
± 8 KV air
± 6 KV contact
± 8 KV air
Floors should be wood,
concrete or ceramic tile. If
the floor is covered with
synthetic material, the
relative humidity should
be at least 30%.