11
5. Cleaning the device
It is recommended that the nebulizer, mouthpiece and mask are
thoroughly cleaned with hot water after each use and cleaned
with a mild detergent after the last treatment of the day. If your
physician or respiratory therapist specifies a different cleaning
procedure, follow their instructions..
Ÿ
Disconnect the air tube, nebulizer, mouthpiece and mask
Ÿ
Prepare one part white vinegar with three parts distilled water.
Make sure mixed solution is enough to submerge the
nebulizer, mouthpiece and mask.
Ÿ
Wash nebulizer, mouthpiece and mask in warm water and a
mild detergent. Then wash them in hot tap water.
Ÿ
Submerge these parts in the vinegar and water solution for
thirty minutes.
Ÿ
Complete rinsing
Filter change
Ÿ
Do not use cotton or any other materials
Ÿ
Do not wash or clean the filter.
Ÿ
Only use filters supplied by your distributor.
Ÿ
Do not operate without a filter.
Ÿ
Change the filter every 30 days or when the filter turn gray.
Filter changing procedure:
Ÿ
Remove the filter cover
Ÿ
Replace the used filter with a new one
Ÿ
Replace the filter cover.
11