Instruction Manual OilGuard 2 W
Mounting
20
14809E/2
4
Mounting
4.1
Location selection
Note the following points for the operating location:
Electrical supply must be ensured.
The water supply must be ensured as described in the technical data.
The water outlet must be unhindered.
The system should not be exposed to direct sunlight during measurement; the meas-
urement can be skewed by excessive external light.
4.2
Mounting the OilGuard 2 W
The dimensional specifications according to the mounting drawings attached in the docu-
mentation ring binder of OilGuard 2 W.
Mount the OilGuard 2 W as follows:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Screw on the mounting bracket at the desired
position. The mounting bracket must be hori-
zontally aligned with a spirit level.
2.
Align the OilGuard 2 W to the two positioning
pins (circles), then screw onto the mounting
bracket.
3.
Remove the optics unit from the OilGuard 2 W
according to Section 9.1.1.