Electrical installation
Instruction Manual ColorPlus 2
13045E/3
41
5.4
Connecting the SICON (M)
DANGER!
Life-threatening voltage inside the instrument.
Connecting electrical lines can be extremely dangerous. Instrument parts may also be dam-
aged. Local regulations for electrical installations must be observed at all times.
5.4.1
Cable cross-sections over longer distances
For connections longer than the standard cable length of 5 meters, an optional junction
box must be connected between the photometer and control unit.
Depending on the cable used (cross-section) and power supply, the distance between
the control unit and the photometer can be calculated according to the table below.
For cable lengths of more than 5 meters, shielded cable must be used.
Figure 10: Arrangement of control units at distances of more than 5 meters
Photometer
Junction box
Control unit
The maximum distance (X) between the control unit and junction box depends on the volt-
age in the SICON and the cable cross-section used:
Cable cross-section
Max. distance
for UV instrument
Max. distance
for VIS instrument
Remarks
[mm
2
]
[m]
[m]
0.14
20
60
0.25
35
100
0.34
50
140
0.50
70
210
Standard version
0.75
100
320
1.00
140
410
1.50
200
590