Servicing
Instruction Manual AquaMaster
12745E/4
95
8.2.6
Cleaning the parts which come into contact with water
The following describes cleaning the parts of the photometer that come into contact with
water:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the sample supply to the photome-
ter.
Section 4.9
2.
Remove the inlet and outlet from the photo-
meter.
3.
Remove the optics unit of the photometer
and fasten onto the docking station.
Section 8.2.1
4.
Loosen the conduit gland (A) and remove the
inlet pipe (B).
5.
Pull off the outlet cone (C) upward.