Signatrol SPYDAQ-1001-T Quick Start Manual Download Page 2

Calibration: 

All calibration data is held within the spYdaq transmitter, this is held together with

the date of the last calibration. The spYdaq transmitter sends the last calibration date every 
time information is sent to the base-station. Signatrol Ltd recommend as 12-month calibration 
interval. Please contact your agent for more information about our calibration services.

Battery: 

All transmitters come with a new battery fitted. Life is dependant on transmit rate and

ambient temperature conditions. Replacements are available from your supplier. 

Estimated battery life

 (Ambient Temperature -10 to +50°C)

RF Power Setting

Transmit Rate
1 minute

Transmit Rate
10 minutes

Transmit Rate
30 minutes

Low

1.4 years

7.9 years

12.6 years

Normal

1.3 years

7.6 years

12.4 years

High

1.2 years

7.4 years

12.0 years

Replacement Battery:

Signatrol Part No:

SPYDAQ-ACC-01

Type

Lithium Thionyl Chloride (Li/SOCI2)

Voltage

3.6V

Capacity

2200mAh

Cell Temp Range

-55 to +85°C

Size

AA

Low Battery Warning: 

The transmitter uses a coulomb counting technique for estimating 

battery life as Lithium Thionyl Chloride batteries cannot be checked for capacity based on 
output voltage. Using this technique the transmitter will indicate a low battery when the total 
cell capacity falls below 10%. The remaining capacity value (0 to 100%) is viewable in 
spYconfig using the USB connection.  The battery capacity calculation is only accurate if the 
battery calculations are reset every time the battery is replaced with a new battery. This can 
be done through spYconfig. Never fit used batteries as the system will not generate the 
correct low battery warning.

Battery

 

Warning

: Fire, explosion and severe burn hazard. Do not recharge, crush, 

disassemble, heat, above 212 °F (100 °C), incinerate, short circuit or expose contents to 
water. Keep battery out of reach of children. Dispose of used batteries promptly and safely.

Operating Ranges:

Channel

Internal Temperature

Sensor Type

NTC Thermistor

Max Temperature Range

-40 to +105°C 

Operating Range within Enclosure

-30 to +55°C

Accuracy

±0.2°C  (-20°C to +50°C)
±0.5°C  (-35°C to +100°C)
±0.8°C  (-40°C to +105°C)

Resolution

0.05°C 

Channel

Internal Humidity (model dependant)

Sensor Type

Digital

Max Humidity Range

0 to 100% RH 

Operating Range within Enclosure

5 to 95% RH

Accuracy Typical

±2%  (0 to 100% RH)

Accuracy Maximum

±2.5%  (0 to 90% RH)
±3.5%  (> 90% RH)

Resolution

0.43% RH 

Repeatability 

0.1% RH + Resolution

Hysteresis

±0.8% @ 25°C

Long Term Drift

< 0.25%/Year (Typical)
 <0.5%/Year (Maximum)

Response Time (t63%) @ 25°C

20 seconds (airflow 1m/Sec)

Channel 

RTD (Resistance)

Sensor Type

External Resistance 3-wire Measurement

Configuration Pt100 EN70651

(fundamental 38.5 ohms)

Range: -200 to +550°C
Accuracy: ±0.1°C ±0.05% Reading  
Thermal Drift Zero: 0.025°C/°C
Thermal Drift Span: TBC
Resolution: 0.05°C
Specifications above exclude sensor errors.

Configuration Pt100 JISC

(fundamental 39.2 ohms)

Range: -100 to +450°C
Accuracy: ±0.1°C ±0.05% Reading  
Thermal Drift Zero: 0.025°C/°C
Thermal Drift Span: TBC
Resolution: 0.05°C
Specifications above exclude sensor errors.

Configuration Ni100

(0.00618 coefficient)

Range: -55 to +175°C
Accuracy: ±0.1°C ±0.05% Reading  
Thermal Drift Zero: 0.015°C/°C
Thermal Drift Span: TBC
Resolution: 0.05°C
Specifications above exclude sensor errors.

Configuration Ni120

(0.00672 coefficient)

Range: -70 to +200°C
Accuracy: ±0.1°C ±0.05% Reading 
Thermal Drift Zero: 0.015°C/°C
Thermal Drift Span: TBC
Resolution: 0.05°C
Specifications above exclude sensor errors.

Configuration Ohms

Range: 1 to 300 ohms
Accuracy:  ±0.04

W  ±0.01

% Reading

Thermal Drift Zero: 8m

W

/°C

Thermal Drift Span: TBC
Resolution: 19m

W

Channel 

Thermocouple (mV)

Sensor Type

External emf

Configuration T/C “T”

Range: -190 to +400
Accuracy: ±0.05% FSD ± CJ
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: 0.1°C
Specifications above exclude sensor errors.

Configuration T/C “K”

Range: -160 to 1200°C
Accuracy: ±0.05% FSD ± CJ
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: 0.1°C
Specifications above exclude sensor errors.

Configuration T/C “J”

Range: -180 to 840°C
Accuracy: ±0.05% FSD ± CJ
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: <0.1°C
Specifications above exclude sensor errors.

Configuration T/C “R”

Range: +200 to 1755°C
Accuracy: ±0.1% FSD ± CJ 
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: <0.5°C (above 400°C)
Specifications above exclude sensor errors.

Configuration T/C “S”

Range: +200 to 1755°C
Accuracy: ±0.1% FSD ± CJ
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: <0.5°C (above 400°C)
Specifications above exclude sensor errors.

Configuration mV

Range:  -8 to +49mV
Accuracy: ±0.025% FSD 
Thermal Drift Zero: 1

m

V/°C

Thermal Drift Span: TBC
Resolution: 4

m

V

CJ (Cold Junction)  See Internal Temperature error

Channel 

D.C Voltage (V)

Sensor Type

External emf

Range: 0 to 10.01V
Accuracy: ±0.025% 
Thermal Drift Zero: TBC
Thermal Drift Span: TBC
Resolution: 0.7mV

Channel 

D.C Current (mA)

Sensor Type

External current

Range: 0 1 to 21mA
Accuracy: ±0.03% 
Thermal Drift Zero: TBC
Thermal Drift Span: TBC
Resolution: 1.3

m

Channel 

RH Temp

Sensor Type

Temperature of external humidity sensor.

Accuracy Typical

±0.2°C   (0 to +90°C)
±0.4°C  (-40 to +100°C)
±0.5°C  (> +100°C)

Accuracy Maximum

±0.4°C  (-40 to +90°C)
±0.7°C  (> +90°C)

Resolution

±0.04°C

Repeatability

0.06°C + Resolution

Long Term Drift

< 0.03°C/Year (Maximum)

SIG-1389-01-01

2 Of 2

CN0111

Reviews: